-29- Q1E-EP1412
5. Cleaning and Disinfection of EZU-PA7C2
EZU-PA7C2 must be reprocessed after each use. Refer to the
reprocessing instruction in this chapter.
WARNINGS
EZU-PA7C2 is delivered unsterile. Prior to the
first use, reprocess it.
The cavities and the spiral spring of the EZU-
P
A7C2
require particular attentions during all
processes. The needle guide bracket must be
manually pre-
c
leaned using ultrasound prior
manual
r
eprocessing as well as automated
reprocessing and must be dismantled into single
parts (shown in Fig. 25).
Transportation
before using
EZU-PA7C2 should be packed i
n
a sterile pouch or
c
ontainer to transport from Central Sterile Supply
Department (CSSD) to an operating room. Be careful
n
ot to damage the sterile pouch or container during
transportation.
Levels of reprocessing requirements: Refer to the chapter 4.