-39- Q1E-EP1412
8. Specifications
8.1 Probe
Type: EUP-B712 Biopsy Probe
Acoustic working Frequency: 3.0MHz
Technology: Convex Array Probe
Dimensions: See Fig. 26.
Weight: Approx. 0.70kg (Including cable and
connector)
Probe materials: Biocompatible allergy free components
Acoustic output: According to IEC60601-2-37 (See Main Unit
manual)
Applicable systems: Depending on production and upgrade
status f
o
r detailed information, contact
a service support.
Classification: MDD classification IIa.
Cleaning:
A
pplicable detergents are listed in the
supplier list
Disinfection: Applicable disinfectants are listed in
the suppliers list
Sterilization: ETO gas sterilization and Plasma
sterilization
Operating conditions:
Ambient temperature: +10 - +40°C
Contact surface temperature
(Temperature of examinee):
max. 42°C
Relative humidity: 30 - 75% (subject to no condensation)
Storage conditions:
Temperature: -10 - +50°C
Relative humidity: 10 - 90% (subject to no condensation)