14
5. LAK2D Series:
5.1 LAK2D can combine uniaxial or bi-axial linear actuators. For the sake of safety, two actuators can’t be
operated simultaneously. Don’t press the “△” key and “▽” key at the sam e tim e.
5.2 The output power of a controller is 108VA (The output voltage is 24VDC.).
5.3 LAK2D is equipped with auto protector to avoid over current caused by overload or other abnormal
situations; two 9V alkaline batteries can be added during temporary interruption to reduce the load.
5.4 LAK2D has no charge function, the battery can only afford the load to be declined urgently under a
powerless state.
5.5 LAK2D can combine LAM3 to save the space and be easily installed (Please refer to the figure below.).
5.6 Fuse specification:2A/20mm(220/230VAC),
3A/20mm(100/110VAC).
6. LAK2J Series(UL approved only DC IN versions):
6.1 LAK2J can combine uniaxial or bi-axial linear actuators.
6.2 A portable controller is designed in this product driven by two 12VDC lead acid batteries with 4.5Ah
without built-in transformer.
6.3 The EMERGENCY button on a panel can disconnect the function of a controller to protect customers
users.
6.4 Users can directly press the UP on DOWN key on a panel without a key box (The button can only be use
to do the up and down control for the first axis).
6.5 Over current protection and slow start are provided to avoid misuse.
6.6 A LAKCH-A charger is required for charging, (For the sake of safety, a controller can’t be used in charging.).
6.7 In 30 seconds after stopping operation, it will automatically become energy-saving mode to increase the
use time. The low-level buzz and LED power indication are provided as well.
6.8 When the controller is idle, please still keep the controller supplied with power (power cable connected),
when the controller is not in use, please press down the emergency button to save battery power.
6.9 Fuse specification:15A/20mm.
6.10 Note: Please charge battery for at least 12 hours before initial use.
6.11 Do not disconnect the cable connector when actuator is running.