Problem: Device is overheated
Device is used with maximum performance long period of time in hot
environment.
Wait until the unit cools down and the error message disappears.
Problem: Suction initiates, but the set negative pressure is not achieved.
Ensure that the treatment cup fits tightly on the area you are attempting to treat. Try to find a
location where you can achieve an airtight connection or replace treatment cup with the size that
fits better.
The filter in the treatment cup may be clogged. Replace the treatment cup.
Verify that the treatment cup is properly connected to the Treatment Head.
Verify that the soft treatment cup cover is properly connected to the treatment cup and intact.
Wear and tear of the treatment cup may result in leaks appearing. Verify and replace the treatment
cup if necessary.