15. EMC
15.1 Guidance and manufacturer’s declaration – electromagnetic emissions
The PhysioTouch® - LymphaTouch® is intended for use in the electromagnetic environment specified
below. The customer or the user of the PhysioTouch® - LymphaTouch® should assure that it is used in
such an environment.
Warning: Use of accessories, transducers and cables other than those specified, with the exception
of transducers and cables sold by HLD Healthy Life Devices as replacement parts for internal
components, may result in increased electromagnetic emissions or decreased immunity of the device
to such emissions.
Warning: PhysioTouch® - LymphaTouch® should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the PhysioTouch® - LymphaTouch® should be
observed to verify normal operation in the configuration in which it will be used.