37
Emission tests according to the test specification IEC 60601-1-2.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The PhysioTouch® - LymphaTouch® uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The PhysioTouch® - LymphaTouch® is suitable for use in
all establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not Applicable
(low input
power)
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies