4 Replacement Kit Instruction 43-DR-33-87 3/05
Electrical Alignment
If the group prompt “LOCKOUT” has been configured for “CALIB”, +VIEW”, or “MAX” selection, you
will not be able to access PEN 1 group prompt for electrical alignment. Set LOCKOUT to “NONE”.
Table 3 Procedure for Setting the Chart Time Line
Step Press Action/Result
1
Open the door, apply power, and wait for
the recorder to run its power up tests.
2
SET
UP
Until you see:
Upper Displa
Lower Display
SET UP
LOCKOUT
3
FUNC
Until you see:
Upper Display
Lower Display
LOCKOUT
One of the following
NONE
CALIB
+ CONF
+ VIEW
MAX
4
or
Until “NONE” is in the lower display.
ATTENTION
Never leave the
“LOCKOUT” selection in “NONE.” Use
“NONE LOCKOUT” only when you intend
to make field adjustments to the
configuration
5
SET
UP
Until you see:
PEN 1
Lower Display
ALIGN
Upper Display
6
FUNC
You will see:
PEN 1 0
Lower Display
ALIGN
Upper Displa
The pen moves to an internal stop then
backs away and stops.
7
or
to move the pen to the zero position on the
chart (inner position of chart).
ATTENTION
You can position the pen
over 1200 steps. Those steps are relative
units only. Step 0.0 is the inner most
position on the chart, and step 1200 is the
highest circumference position.