HP 3457A - 5-23. DC Volts Gain Calibration - Front Terminals

HP 3457A
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
3457A
4.
Press
the
ENT
key.
The
display
will
show
CALIBRATING
while
the
calibration
routine
is
running.
eS
d,
Repeat
steps
"cl
"
through
"
c4
"
for
each
of
the
remaining
voltage
ranges.
e.
Remove
the
short
from
the
front
panel
HI
and
LO
Input
Terminals.
5-23.
DC
Volts
Gain
Calibration
-
Front
Terminals
5-24.
Equipment
Required.
A
DC
Voltage
Source
capable
of
providing
30
mV
(+.0055%),
300
mV
(+.0012%),
3
V
(+.0006%),
30
V
(+.0013%)
and
300
V
(+.0017%)
is
required
for
the
following
procedure.
a.
Set
the
HP
3457A
to
the
300
volt
range.
(Use
the
|
scroll
key
to
select
the
300
V
range).
b.
Connect
the
Voltage
Source
to
the
HP
3457A
front
panel
HI
and
LO
input
terminals
and
set
its
output
to
300
V.
c.
Use
the
following
steps
to
calibrate
the
DC
Voltage
Gain
on
the
300
V,
30
V,
3
V,
300
mV
and
30
mV
ranges
beginning
with
the
300
V
range
and
ending
on
the
30
mV
range.
1,
Set
the
3457A
to
the
appropriate
voltage
range.
(Use
the
{
or
|
scroll
key
to
select
the
proper
voltage
range).
2,
Set
the
voltage
source
to
an
output
which
will
provide
a
full-scale
reading
on
the
HP
3457A.
3.
Access
the
Calibration
Command.
(Press
the
Blue
SHIFT
key,
then
the
OFFSET
COMP
/
C
configura-
tion
key),
The
display
will
show
CAL.
r
4,
Enter
the
value
of
the
input
voltage
(in
volts).
(NOTE:
If
the
security
feature
has
not
been
disabled,
(set
to
0)
it
will
be
necessary
to
enter
the
input
value,
the
delimiter
(
,
)
and
the
security
code).
5.
Press
the
ENT
key.
The
display
will
show
CALIBRATING
while
the
calibration
routine
is
running,
d.
Repeat
steps
"cl"
through
"c5
"
for
each
of
the
remaining
voltage
ranges.
e.
Set
the
output
of
the
Voltage
Source
to
3
volts.
5-25.
DC
Volts
Linearity
Calibration
-
Front
Terminals
5-26.
Equipment
Required.
A
DC
Voltage
Source
capable
of
providing
3
V
(+.0006%)
is
required
to
calibrate
linearity
of
the
DC
Voltage
Function.
a.
Set
the
HP
3457A
to
the
3
volt
range.
b.
Set
the
Voltage
Source
for
an
output
voltage
of
3
Vdc.
c.
Reverse
the
leads
at
the
3457A
input
terminals
to
provide
a
negative
input
voltage.
d.
Access
the
Calibration
Command.
(Press
the
Blue
SHIFT
key
then
the
OFFSET
COMP
/
C
configuration
key).
The
display
will
will
show
CAL.
#