EasyManua.ls Logo

HP 3457A Service Manual

HP 3457A
273 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
HP
3457A
Multimeter
HP
3457A
Service
Manual
|
@
HEWLETT
PACKARD
Copyright
©
Hewlett-Packard
Company,
1988
Manual
Part
Number:
03457-90012
Edition
3
Microfiche
Part
Number:
03457-99012
E0288

Other manuals for HP 3457A

Question and Answer IconNeed help?

Do you have a question about the HP 3457A and is the answer not in the manual?

HP 3457A Specifications

General IconGeneral
Model3457A
CategoryMultimeter
TypeDigital Multimeter
DC Voltage Range100 mV to 1000 V
AC Voltage Frequency Range10 Hz to 1 MHz
ManufacturerHewlett-Packard
InterfaceIEEE-488

Summary

Printing History

SAFETY SUMMARY

Operating and Safety Symbols

Describes symbols used on products and in manuals, including their meanings and applications.

SECTION I GENERAL INFORMATION

1-1. INTRODUCTION

Provides an overview of the manual's content, covering installation, operation, performance testing, calibration, and service.

1-3. MANUAL DESCRIPTION

Details the manual's layout and content, outlining the structure of its various sections.

1-5. INSTRUMENT DESCRIPTION

Describes the HP 3457A multimeter's capabilities, including measurement functions and features like high-speed applications.

1-7. SAFETY CONSIDERATIONS

Highlights critical safety precautions and warnings for operating and servicing the instrument.

1-13. SPECIFICATIONS

Lists the performance standards and specifications that the instrument is guaranteed to meet.

1-15. OPTIONS AND ACCESSORIES

Details the available options and accessories that can be used with the HP 3457A multimeter.

1-17. TEST EQUIPMENT

Lists the necessary test equipment required for calibrating and repairing the HP 3457A multimeter.

SECTION II INSTALLATION

2-1. INTRODUCTION

Introduces installation procedures, including initial inspection, power requirements, and repackaging instructions.

2-3. INITIAL INSPECTION

Provides guidelines for inspecting the instrument for mechanical and electrical condition upon receipt.

2-5. PREPARATION FOR USE

Covers essential preparation steps before using the instrument, including power and fuse selection.

2-8. Line Voltage Selection

Explains how to select the appropriate line voltage switches and fuse requirements for different power line voltages.

2-12. Grounding Requirements

Details the grounding requirements to protect personnel from shock hazards by grounding the instrument cabinet.

2-18 INTERFACE CONNECTIONS

Describes interface connections, focusing on the HP-IB interface for system integration.

2-19. Control Interface

Explains the control interface, specifically the HP-IB compatibility and system connection methods.

2-21. HP-IB ADDRESS SELECTION

Details how to set the HP-IB address, both from the front panel and via the interface bus.

2-23. Changing the Address from the Front Panel

Provides step-by-step instructions for changing the HP-IB address using the front panel controls.

2-24. Changing the Address from the Controller

Explains how to change the HP-IB address of the instrument using an external controller.

2-25. INSTALLATION OF OPTIONS

Covers the installation of optional plug-in assemblies, including configuration and wiring.

2-28. Rear Terminal Assembly Removal

Provides instructions for removing the rear terminal assembly prior to installing plug-in assemblies.

2-30. Plug-In Assembly Installation

Details the procedure for installing plug-in assemblies, including connecting the wiring harness.

2-32. Plug-In Assembly Configuration

Explains how to configure plug-in assemblies, dependent on the specific assembly and user application.

2-34. HP 44491A Configuration

Describes the configuration options for the HP 44491A Armature Relay Assembly.

2-36. HP 44492A Configuration

Details the configuration options for the HP 44492A Reed Relay Multiplexer Assembly.

2-38. Wiring Block Preparation

Outlines the procedures for preparing the wiring block for plug-in assembly installation.

SECTION III OPERATION

3-1. INTRODUCTION

Introduces the manual operation section, focusing on front panel operations for service personnel.

3-3. GENERAL INFORMATION

Provides general information on operating the HP 3457A multimeter.

3-11. Voltage Measurements (Manual Operation)

Explains how to perform manual voltage measurements, including function and range selection.

3-12. Function Selection

Details how to select measurement functions like DC Volts, AC Volts, and DC Coupled AC Volts.

3-13. Range Selection

Describes how to select measurement ranges, including Auto-Range and Manual Ranging features.

3-14. Measurement Resolution

Explains how to change the measurement resolution by selecting the number of digits displayed.

3-15. Integration Time

Details how to change the integration time to reduce noise or permit faster readings.

3-18. Resistance Measurements (Manual Operation)

Covers manual resistance measurements, including function and range selection.

3-19. Function Selection

Explains how to select measurement functions for resistance, including 2-Wire and 4-Wire Ohms.

3-20. Range Selection

Describes how to select resistance measurement ranges, including Auto-Range and Manual Ranging.

3-21. Measurement Resolution

Explains how to change measurement resolution by selecting the number of digits displayed.

3-25. Current Measurements (Manual Operation)

Covers manual current measurements, including function and range selection.

3-26. Function Selection

Details how to select measurement functions for current: DC Current, AC Current, and DC Coupled AC Current.

3-32. Frequency/Period Measurements (Manual Operation)

Explains manual measurements for frequency and period, including function and range selection.

3-33. Function Selection

Details how to select measurement functions for frequency and period.

3-44. Scanning

Describes how to set up automatic scanning to take and store measurements from multiple channels.

3-50. Scanning

Details the process for setting up automatic scanning to measure inputs and store results.

FUNCTION KEYS

Explains the function keys used to select the type of measurement for the instrument.

ANNUNCIATORS

Describes the annunciators across the display that indicate the instrument's operating conditions.

DISPLAY

Details the 12-character alphanumeric LCD display used for readings and messages.

CONFIGURATION KEYS

Explains how to set instrument parameters and configure operating settings using configuration keys.

INPUT TERMINALS

Describes the input terminals for voltage, resistance, and current measurements.

SECTION IV PERFORMANCE TESTS

4-1. INTRODUCTION

Introduces performance tests designed to verify the accuracy of the HP 3457A multimeter.

4-3. EQUIPMENT REQUIRED

Lists the equipment required for performing performance and operational verification tests.

DC Voltage Performance Tests

Details the procedures for testing the DC voltage measurement capabilities of the instrument.

4-14. DC Voltage Function - Offset Test

Describes the procedure for testing the DC voltage function's offset, requiring a low thermal short.

4-16. DC Voltage Function - Gain Test

Explains the procedure for testing the DC voltage function's gain, requiring a DC voltage standard.

AC Voltage Performance Tests

Outlines the procedures for testing the AC voltage measurement capabilities of the instrument.

4-20. AC Voltage Function - Gain Test

Details the procedure for testing the AC voltage function's gain, requiring an AC signal source.

4-22. AC Voltage Function - Frequency Response Test

Explains the procedure for testing the AC voltage function's frequency response using an AC signal source.

DC Current Performance Tests

Details the procedures for testing the DC current measurement capabilities of the instrument.

4-24. DC Current Function - Offset Test

Describes the procedure for testing the DC current function's offset, requiring no specific test equipment.

4-26. DC Current Function - Gain Test

Explains the procedure for testing the DC current function's gain, requiring a DC current source.

AC Current Performance Tests

Outlines the procedures for testing the AC current measurement capabilities of the instrument.

4-28. AC Current Function - Gain Test

Details the procedure for testing the AC current function's gain, requiring an AC current source.

2-Wire Ohms Performance Tests

Covers the procedures for testing the 2-Wire Ohms measurement capabilities.

4-30. 2-Wire Ohms Function - Offset Test

Describes the procedure for testing the 2-Wire Ohms function's offset, requiring a low thermal short.

4-32. 2-Wire Ohms Function - Gain Test

Explains the procedure for testing the 2-Wire Ohms function's gain, requiring resistance standards.

4-Wire Ohms Performance Tests

Details the procedures for testing the 4-Wire Ohms measurement capabilities.

4-34. 4-Wire Ohms Function - Offset Test

Describes the procedure for testing the 4-Wire Ohms function's offset, requiring a low thermal short.

4-36. 4-Wire Ohms Function - Gain Test

Explains the procedure for testing the 4-Wire Ohms function's gain, requiring resistance standards.

Frequency Counter Performance Tests

Covers the procedures for testing the frequency counter measurement capabilities.

4-38. Frequency Counter - Accuracy Test

Details the procedure for testing the frequency counter's accuracy using a frequency source.

Rear Input Performance Tests (Standard Instrument)

Outlines performance tests for the instrument's rear input terminals.

4-41. DC Voltage Function - Rear Terminal Offset Test

Describes testing the DC voltage function's rear terminal offset, requiring a low thermal short.

4-43. DC Current Function - Rear Terminal Offset Test

Details testing the DC current function's rear terminal offset, requiring no specific test equipment.

4-45. 2-Wire Ohms Function - Rear Terminal Offset Test

Describes testing the 2-Wire Ohms function's rear terminal offset, requiring a low thermal short.

4-47. 4-Wire Ohms Function - Offset Test

Explains testing the 4-Wire Ohms function's offset, requiring a low thermal short.

Rear Input Performance Tests (44491A General Purpose Relay Assy)

Outlines performance tests for the 44491A General Purpose Relay Assembly rear inputs.

4-50. DC Voltage Function - 44491A Offset Test

Details testing the DC voltage function's offset for the 44491A assembly, requiring shorting wires.

4-52. 2-Wire Ohms Function - 44491A Offset Test Limits

Provides limits for testing the 2-Wire Ohms function's offset on the 44491A assembly.

4-54. 4-Wire Ohms Function - 44491A Offset Test Limits

Lists limits for testing the 4-Wire Ohms function's offset on the 44491A assembly.

4-56. DC Current Function - 44491A Offset Test Limits

Provides limits for testing the DC current function's offset on the 44491A assembly.

Rear Input Performance Tests (44492A 10 Channel Multiplex Assy)

Outlines performance tests for the 44492A 10 Channel Multiplexer Assembly rear inputs.

4-59. DC Voltage Function - 44492A Offset Test

Details testing the DC voltage function's offset for the 44492A assembly, requiring shorting wires.

4-61. 2-Wire Ohms Function - 44492A Offset Test Limits

Lists limits for testing the 2-Wire Ohms function's offset on the 44492A assembly.

Operational Verification Tests

DC Voltage Operational Verification Tests

Covers verification tests for DC voltage measurements, including offset and gain.

4-66. DC Voltage Function - Offset Test

Details the procedure for testing the DC voltage function's offset during operational verification.

4-68. DC Voltage Function - Gain Test

Explains the procedure for testing the DC voltage function's gain during operational verification.

AC Voltage Operational Verification Tests

Outlines verification tests for AC voltage measurements, including gain and frequency response.

4-70. AC Voltage Function - Gain Test

Details the procedure for testing the AC voltage function's gain during operational verification.

4-72. AC Voltage Function - Frequency Response Test

Explains the procedure for testing the AC voltage function's frequency response during verification.

DC Current Operational Verification Tests

Covers verification tests for DC current measurements, including offset and gain.

4-74. DC Current Function - Offset Test

Describes the procedure for testing the DC current function's offset during operational verification.

4-76. DC Current Function - Gain Test

Explains the procedure for testing the DC current function's gain during operational verification.

AC Current Operational Verification Tests

Outlines verification tests for AC current measurements, focusing on gain.

4-78. AC Current Function - Gain Test

Details the procedure for testing the AC current function's gain during operational verification.

2-Wire Ohms Operational Verification Tests

Covers verification tests for 2-Wire Ohms measurements, including offset and gain.

4-80. 2-Wire Ohms Function - Offset Test

Describes the procedure for testing the 2-Wire Ohms function's offset during verification.

4-82. 2-Wire Ohms Function - Gain Test

Explains the procedure for testing the 2-Wire Ohms function's gain during verification.

4-Wire Ohms Operational Verification Tests

Outlines verification tests for 4-Wire Ohms measurements, including offset and gain.

4-84. 4-Wire Ohms Function - Offset Test

Details the procedure for testing the 4-Wire Ohms function's offset during verification.

4-86. 4-Wire Ohms Function - Gain Test

Explains the procedure for testing the 4-Wire Ohms function's gain during verification.

Frequency Counter Operational Verification Tests

Covers verification tests for the frequency counter measurement capabilities.

4-88. Frequency Counter - Accuracy Test

Details the procedure for testing the frequency counter's accuracy during verification.

Rear Input Performance Tests (Standard Instrument)

Outlines performance tests for the standard instrument's rear input terminals.

4-91. DC Voltage Function - Rear Terminal Offset Test

Describes testing the DC voltage function's rear terminal offset during verification.

4-93. DC Current Function - Rear Terminal Offset Test

Details testing the DC current function's rear terminal offset during verification.

4-95. 2-Wire Ohms Function - Rear Terminal Offset Test

Describes testing the 2-Wire Ohms function's rear terminal offset during verification.

4-97. 4-Wire Ohms Function - Offset Test

Explains testing the 4-Wire Ohms function's offset during verification.

SECTION V CALIBRATION

5-1. INTRODUCTION

Introduces the manual calibration procedures for the HP 3457A multimeter.

5-4. CALIBRATION SECURITY

Explains the calibration security feature that prevents unauthorized calibration.

5-8. Changing the Security Code

Provides the procedure for changing the security code to prevent accidental calibration.

5-13. PRELIMINARY CALIBRATION PROCEDURES

Outlines essential steps to perform before calibrating the HP 3457A multimeter.

5-15. Input Amplifier Offset Adjustment

Details the adjustment procedure for the input amplifier offset, required for specific self-test errors.

5-17. AC Convertor Frequency Response Adjustment

Explains the adjustment for AC convertor frequency response, needed for specific self-test errors.

5-19. CALIBRATION PROCEDURES - FRONT PANEL INPUTS

Covers calibration procedures for offset and gain using front panel inputs.

5-21. DC Volts Offset Calibration - Front Terminals

Details the procedure for calibrating the DC volts offset using front panel terminals.

5-23. DC Volts Gain Calibration - Front Terminals

Explains the procedure for calibrating the DC volts gain using front panel terminals.

5-27. Two-Wire Ohms Offset Calibration - Front Terminals

Details the procedure for calibrating the 2-Wire Ohms offset using front panel terminals.

5-29. Two-Wire Ohms Gain Calibration - Front Terminals

Explains the procedure for calibrating the 2-Wire Ohms gain using front panel terminals.

5-31. Four-Wire Ohms Offset Calibration - Front Terminals

Details the procedure for calibrating the 4-Wire Ohms offset using front panel terminals.

5-33. Four-Wire Ohms Gain Calibration - Front Terminals

Explains the procedure for calibrating the 4-Wire Ohms gain using front panel terminals.

5-35. DC Current Offset Calibration - Front Terminals

Details the procedure for calibrating the DC current offset using front panel terminals.

5-37. DC Current Gain Calibration - Front Terminals

Explains the procedure for calibrating the DC current gain using front panel terminals.

5-39. AC Volts Offset & Gain Calibration - Front Terminals

Details calibrating AC volts offset and gain using front panel terminals.

5-41. AC Current Calibration - Front Terminals

Explains calibrating AC current offset and gain using front panel terminals.

5-43. Frequency Calibration - Front Terminals

Details calibrating the frequency function using front panel terminals.

5-45. CALIBRATION PROCEDURES - REAR PANEL INPUTS

Covers calibration procedures for the instrument's rear panel inputs.

5-47. DC Volts Offset Calibration - Rear Terminals (Standard Instrument)

Details calibrating DC volts offset for rear terminals on the standard instrument.

5-49. Two-Wire Ohms Offset Calibration - Rear Terminals (Standard Instrument)

Explains calibrating 2-Wire Ohms offset for rear terminals on the standard instrument.

5-51. Four-Wire Ohms Offset Calibration - Rear Terminals (Standard Instrument)

Details calibrating 4-Wire Ohms offset for rear terminals on the standard instrument.

5-53. DC Current Offset Calibration - Rear Terminals (Standard Instrument)

Explains calibrating DC current offset for rear terminals on the standard instrument.

5-55. CALIBRATION PROCEDURES - HP 44491A GENERAL PURPOSE RELAY ASSEMBLY

Covers calibration procedures specific to the HP 44491A General Purpose Relay Assembly.

5-56. Wiring Block Preparation

Outlines the procedures for preparing the wiring block before calibration for the 44491A assembly.

5-58. DC Volts Offset Calibration - HP 44491A

Details calibrating DC volts offset for the HP 44491A assembly.

5-59. Two-Wire Ohms Offset Calibration - HP 44491A

Explains calibrating 2-Wire Ohms offset for the HP 44491A assembly.

5-60. Four-Wire Ohms Offset Calibration - HP 44491A

Details calibrating 4-Wire Ohms offset for the HP 44491A assembly.

5-61. DC Current Offset Calibration - HP 44491A

Explains calibrating DC current offset for the HP 44491A assembly.

5-62. CALIBRATION PROCEDURES - HP 44492A 10 CHANNEL MULTIPLEXER ASSEMBLY

Covers calibration procedures for the HP 44492A 10 Channel Multiplexer Assembly.

5-63. Wiring Block Preparation

Outlines procedures for preparing the wiring block before calibration for the 44492A assembly.

5-65. DC Volts Offset Calibration - HP 44492A

Details calibrating DC volts offset for the HP 44492A assembly.

5-64. Two-Wire Ohms Offset Calibration - HP 44492A

Explains calibrating 2-Wire Ohms offset for the HP 44492A assembly.

SECTION VI REPLACEABLE PARTS

6-1. INTRODUCTION

Introduces the section on ordering replacement parts for the instrument.

6-3. ORDERING INFORMATION

Provides information on how to order listed and non-listed parts, including direct mail order system.

6-15. DISASSEMBLY/ASSEMBLY PROCEDURES

Explains procedures for removing and replacing cabinet parts and circuit assemblies.

6-17. Rear Terminal Assembly Removal

Details the steps for removing the rear terminal assembly from the instrument.

6-19. Optional Plug-In Assembly Removal

Provides instructions for removing optional plug-in assemblies from the instrument.

6-21. Top Cover Removal

Explains the procedure for removing the top cover of the instrument.

6-23. Display Assembly Removal

Details the steps for removing the display assembly from the instrument.

6-25. Digital (A1 or A11) Assembly Removal

Provides instructions for removing the digital assembly (A1 or A11) from the instrument.

6-27. Plug-In Support Assembly Removal

Explains how to remove the plug-in support assembly from the instrument.

6-29. AC Convertor (A3) Assembly Removal

Details the procedure for removing the AC convertor (A3) assembly.

6-31. Analog (A2) Assembly Removal

Provides instructions for removing the analog assembly (A2) from the instrument.

6-33. Front Panel Removal

Explains the procedure for removing the front panel assembly from the instrument.

SECTION VII MANUAL CHANGES

7-1. INTRODUCTION

Introduces the section on manual changes, noting its purpose for adapting the manual to older instruments.

SECTION VIII SERVICE

8-1. INTRODUCTION

Introduces the service section, covering troubleshooting, repair, safety, and theory of operation.

8-2. SAFETY CONSIDERATIONS

Highlights safety considerations for servicing the instrument, emphasizing adherence to manual warnings.

8-3. STATIC HANDLING

Provides essential guidelines for handling components safely to prevent electrostatic discharge (ESD) damage.

8-4. RECOMMENDED TEST EQUIPMENT

Lists the recommended test equipment required for testing and repairing the HP 3457A.

8-5. GENERAL BLOCK DIAGRAM THEORY OF OPERATION

Provides a simplified explanation of the HP 3457A's circuit operation, divided into sections.

8-18. DETAILED THEORY OF OPERATION

Offers a detailed description of the HP 3457A's operating circuitry, explaining each section.

8-19. Input Circuitry

Explains the purpose and operation of the input circuitry in conditioning input signals for the A/D converter.

8-20. Over Voltage/Over Current Protection

Details the circuitry that protects the instrument from over voltage and over current conditions.

8-21. Input Switching

Describes the input switching circuitry, consisting of relays and MOSFET switches for signal routing.

8-23. Offset DAC

Explains the offset DAC section responsible for providing bipolar voltage output for calibration.

8-24. Pre-charge Stage and Pre-charge Offset Adjust

Details the pre-charge stage for input capacitance and the offset adjustment procedure.

8-25. Input Amplifier

Describes the input amplifier's purpose in conditioning input signals and its block structure.

8-26. Ohms Current Source

Explains the purpose of the Ohms Current Source in providing stable current for resistance measurements.

8-27. Ohms Voltage Reference

Details the Ohms Voltage Reference circuit used for precise and stable ohms function measurements.

8-28. Switch Amplifier

Describes the switch amplifier used to enable range resistors for output current value selection.

8-29. Output FET Gate Bias Amplifier

Explains the output FET Gate Bias Amplifier's role in the voltage-to-current converter.

8-30. Offset Compensation Switching (OCOMP command)

Details the offset compensation technique used to remove errors in ohms measurements.

8-33. A/D Analog

Describes the A/D Analog section, including the Integrator, Slope Amplifier, and Auto-Zero Loop.

A/D Integrator (U512 and associated circuitry)

Explains the function of the integrator in accumulating charge for measurement and nulling.

Slope Amplifier (U513)

Describes the slope amplifier's role in amplifying smaller slopes for improved resolution.

DA Compensation (U514 and Q512)

Details the DA compensation circuit used to minimize linearity errors from integrator capacitor DA.

A/D Analog Auto-Zero Loop

Explains the function of the analog auto-zero loop in achieving precision A/D performance.

8-34. Voltage Reference

Describes the precision voltage reference assembly used for maintaining long-term accuracy.

8-35. A/D Controller

Discusses the A/D Controller circuitry, its functions, and divisions.

8-36. A/D Control

Explains how the A/D input signal is integrated and controlled during conversion cycles.

8-37. Frequency Counter

Describes the AC assembly's front end for frequency measurements and its ranging capability.

8-38. Floating Common Microprocessor

Explains the role of the floating common MPU in controlling the Analog Section's measurement dynamics.

8-39. Cross-Isolation Link

Details the cross-isolation link for precision floating measurements, achieved via opto-isolators.

8-40. Main Controller (03457-66501)

Discusses the main controller assembly's functions in managing measurements and communication.

8-41. Reading instructions from ROM U503

Explains the procedure for reading instructions from ROM U503 via the microprocessor.

8-43. Reading information from RAM U506

Details the procedure for reading information from RAM U506 via the microprocessor.

8-44. Reading information from RAM U511

Explains how to read information from RAM U511 via the microprocessor.

8-45. Writing information to RAM U506

Details the procedure for storing information into RAM U506 via the microprocessor.

8-46. Writing to the unprotected section of RAM U511

Explains writing to the unprotected section of RAM U511 via the microprocessor.

8-53. Writing to the protected section of RAM U603

Details writing to the protected section of RAM U603, including CAL-RAM lock operation.

8-54. Reset Circuit Operation

Explains the purpose and conditions for the microprocessor reset circuit operation.

8-55. AC to DC Converter

Describes the AC to DC converter's function in converting AC signals to DC voltage for measurement.

8-56. Input Coupling

Explains how the AC input signal is routed to the AC to DC Converter assembly.

8-57. Attenuators

Details the two input attenuators used to condition the AC input signal for measurement.

8-58. Amplifier Circuit

Describes the AC converter amplifier's two-stage design for providing input voltage to the RMS converter.

8-59. RMS Converter

Explains the RMS converter's function in converting AC signals to DC voltage with high performance.

8-60. DC Offset Correction

Details DC offset correction for ACV function measurements to reduce measurement errors.

8-61. Frequency Comparator

Describes the comparator selected for frequency or period functions to direct the AC signal.

8-62. Automatic Calibration

Explains the automatic calibration function (ACAL) for compensating DC offsets and frequency response.

Auto-zero

Describes the auto-zero process performed at the beginning of each automatic calibration routine.

Frequency Response Compensation

Details frequency response compensation for AC measurements using RC attenuators.

Ohms Precharge Amplifier Auto Calibration

Explains the auto-calibration procedure for the Ohms Precharge Amplifier, similar to Self Test 9.

Extended Ohms Range Auto Calibration

Describes the auto-calibration for extended ohms ranges, focusing on measurement and checking values.

8-63. 44491A General Purpose Relay Card

Provides discussion on the 44491A General Purpose Relay Card's functions and circuitry.

8-64. Control Circuitry

Explains the control circuitry for the relays, driven by shift registers and logic gates.

8-65. Drive Circuitry

Details the matrix scheme of row and column drivers used to control the card's relays.

8-66. Contact Circuitry

Describes the two types of relays used in the 44491A and their contact configurations.

8-67. Operation of Channel 0

Explains the operation of channel 0, including relay configuration and signal paths.

8-68. Card Configuration

Describes how users can configure different channels and components on the circuit board.

8-69. 44492A 10 Channel Reed Multiplexer

Provides discussion on the 44492A 10 Channel Reed Multiplexer's capabilities and ID line.

8-70. Closing a Channel

Details the procedure for closing a channel on the instrument's inguard controller.

8-71. TROUBLESHOOTING

Provides information to assist in isolating failures to the component level on the multimeter module.

8-72. General Information

Discusses special configurations like Self Test, Diagnostic, and Autocal modes.

8-73. A1 Assembly Failure Indications

Describes how to interpret front panel display messages indicating Al assembly faults.

8-74. A2/A3 Assembly Failure Indications

Explains how failures in A2/A3 assemblies are detected via TEST FAILED and error annunciator.

8-75. Self Tests

Details the two categories of self tests: power-on and keyboard executable.

8-76. Power-on Self Test

Describes the abbreviated self test executed at instrument power-on for fundamental hardware checks.

8-77. Self Test (TEST)

Explains the execution of 7 additional self tests using the TEST command.

8-80. Updated Self Tests 0, 1, and 2 (TEST)

Details updated self tests for the Main Controller assembly, including ROM and RAM checks.

8-81. Autocal (ACAL)

Explains the Autocal routine for calibrating AC amplifier DC offset and Attenuator Flatness.

AC Amplifier DC Offset Auto Calibration

Details the auto-calibration procedure for the AC amplifier's DC offset.

AC Amplifier Attenuator Flatness Auto Calibration

Explains the auto-calibration for AC amplifier's attenuator flatness.

Ohms Precharge Amplifier Auto Calibration

Describes the auto-calibration for the Ohms Precharge Amplifier, similar to Self Test 9.

Extended Ohms Range Auto Calibration

Explains the auto-calibration for extended ohms ranges, focusing on measurement and checking values.

8-82. Plug-in Cards

Provides procedures for verifying faulty operation or finding failed components on plug-in cards.

8-83. 44491A

Troubleshooting steps for when the instrument does not work with the 44491A card installed.

8-84. 44492A

Troubleshooting steps for when the instrument does not work with the 44492A card installed.

8-85. Replacement Procedures

References Section VI for procedures on removing and replacing PC assemblies.

Errata

HP References in this Manual

Notes references to HP, now Agilent Technologies, and potential manual copy inconsistencies.

About this Manual

Provides information about the manual's availability on the Agilent website and potential quality issues.

Support for Your Product

Informs users that Agilent no longer sells or supports the product, directing them to the website for information.

Related product manuals