SECTION
IV
PERFORMANCE
TESTS
4-1.
INTRODUCTION
4-2,
Section
IV
contains
Performance
Tests
designed
to
verify
the
accuracy
of
the
HP
3457A
Multimeter.
Accuracy
specifications
are
listed
in
Table
1-1
of
this
manual.
This
section
also
contains
Operational
Verification
Tests
which
provide
a
more
rapid
method
of
testing
the
operation
and
accuracy
of
the
unit.
The
Operational
Verification
Tests
are
designed
to
provide
a
90%
confidence
that
the
HP
3457A
is
opera-
tional
and
meets
specifications.
Both
the
Performance
Tests
and
Operational
Verification
Tests
can
be
per-
formed
without
access
to
the
interior
of
the
instrument.
4-3.
EQUIPMENT
REQUIRED
4-4,
Equipment
required
for the
performance
tests
and
operational
verification
tests
are
listed
in
the
Recommend
Test
Equipment
table
in
Section
I
of
this
manual.
Equipment
other
than
that
recommended
may
be
used
as
long
as
the
critical
specifications
are
met
4-5.
TEST
RECORD
4-6,
Results
of
the
performance
or
operational
verification
tests
may
be
tabulated
on
the
Test
Record
lo-
cated
at
the
end
of
the
respective
procedures.
The
Test
Record
lists
all
of
the
tested
specifications
and
their
acceptable
limits.
It
is
suggested
that
the
performance
tests
or
operational
verification
tests
be
performed
and
the
results
tabulated
when
the
instrument
is
received.
These
results
can
be
used
for
comparison
with
periodic
calibration
results.
4-7.
CALIBRATION
CYCLE
4-8.
This
instrument
requires
periodic
performance
verification.
The
frequency
at
which
the
instrument
should
be
tested
is
dependent
upon
its
usage
and
the
environmental
operating
conditions.
To
maintain
24
hour
specifications,
the
instrument
should
be
checked
daily;
to
maintain
90
day
specifications,
the
instru-
ment
should
be
checked
at
90
day
intervals.
It
is
suggested
that
the
performance
test
be
performed
at
90
day
intervals
for
normal
operation.
4-9.
TEST
CONSIDERATIONS
4-10.
General
4-11,
Because
the
HP
3457A
is
capable
of
making
high
accuracy
measurements,
certain
requirements
need
to
be
made.
For
example,
standards
being
used
to
test
accuracy
should
not
introduce
any
significant
uncer-
tainties
in
the
performance
tests.
A
standard
which
is
ten
times
more
accurate
than
the
HP
3457A
nearly
eliminates
uncertainties.
In
most
cases,
standards
with
these
accuracies
are
not
readily
available,
therefore,
a
compromise
is
necessary.
A
primary
in
house
standard,
one
which
has
been
certified
by
the
National
Bureau
of
Standards
(NBS)
and
can
supply
the
necessary
outputs,
is
recommended.
If
a
primary
standard
is
not
available,
one
of
the
following
may
be
appropriate:
a.
Use
a
standard
which
is
three
or
four
times
more
accurate
than
the
HP
3457A
specifications
to
be
tested.
Keep
in
mind,
however,
the
potential
uncertainties
these
standards
may
introduce.
b,
Use
a
highly
stable
calibrated
standard
and add
the
correction
factors
(usually
supplied
on
the
calibration
charts)
to
the
test
reading.
c.
Send
the
HP
3457A
to
an
HP
Service
Center
or
other
NBS-certified
standards
facility
for
calibration.
4-1