HP 3457A - DC Voltage Performance Tests; 4-14. DC Voltage Function - Offset Test

HP 3457A
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance
Tests
3457A
4-12.
PERFORMANCE
TESTS
The
Performance
Tests
are
divided
into
6
main
tests,
DC
Volts,
AC
Volts,
DC
Current,
AC
Current,
Ohms,
and
Frequency.
Each
test
can
be
performed
independently
of
the
others.
The
steps
within
each
test,
however,
should
be
performed
in
order.
VOLTAGE
SOURCE
Figure
4-1.
DCV
Performance
Test
Connections
The
temperature
of
the
environment
where
these
tests
are
to
be
performed
must
be
»
within
C
of
the
temperature
where
the
instrument
was
calibrated.
The
instrument
was
calibrated
at
the
factory
in
an
area
with
a
temperature
of
24°
C
(+
C).
4-13,
Preliminary
Steps
1.
Turn
the
instrument
ON
and
allow
a
one
hour
warm-up
period.
2.
Be
certain
all
external
inputs
are
disconnected
from
the
HP
3457A.
Run
the
TEST
routine.
The
test
result
must
read
"SELF
TEST
OK".
If
the
display
shows
"TEST
FAILED",
the
instrument
probably
needs
repair.
3.
Run
the
AUTO
CAL
routine.
(Press
the
AUTO
CAL
key,
enter
the
number
|
and
press
the
ENT
key).
DC
Voltage
Performance
Tests
4-14.
DC
Voltage
Function
-
Offset
Test
4-15.
Equipment
Required.
A
low
thermal
short
(copper
wire)
is
required
for
this
procedure.
1.
Connect
a
low
thermal
short
across
the
Front
Panel
HI
and
LO
Input
Terminals.
2.
Set
the
HP
3457A
to
the
DC
Voltage
function
(DCV)
and
the
number
of
digits
displayed
(DIGITS
DISP)
to
six.
&