HP 3457A - 4-Wire Ohms Operational Verification Tests; 4-84. 4-Wire Ohms Function - Offset Test

HP 3457A
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operational
Verification
Tests
2-Wire
Ohms
Operational
Verification
Tests
Cont'd
34578 3457A
|
3457A
90
day
Limits
1
year
Limits
Input
Range
|
Set
Up
High
|
Low
High
Low
3
Kohm|
3
Kohm|
OHM
3.000312
Kohm
2.999688
Kohm
3.000357
Kohm
2.999643
Kohm
30
Kohm]
30
Kohm|
OHM
30.00132
Kohm
29.99868
Kohm
30.00177
Kohm
29.99823
Kohm
300
Kohm}
300
Kohm|
OHM
300.0130
Kohm
299.9870
Kohm
300.0160
Kohm
299.9840
Kohm
Table
4-26.
2-Wire
Ohms
Gain
Test
Limits
4-Wire
Ohms
Operational
Verification
Tests
_—
-
——s
SENSE
INPUT
lawn
tne,
—_—
natu
~
PATH.
RECALL
STORE)
_
sng
MEAN
A
SOEV
RLNSAMP
|
ERROR.
=
a
a
Ta
|
C4 co
cee
roe
fas)
Figure
4-17.
4-Wire
Ohms
Operational
Verification
Test
Connections
4-84.
4-Wire
Ohms
Function
-
Offset
Test
4-85.
Equipment
Required.
A
low
thermal
short
(copper
wire)
is
required
for
this
procedure.
1.
Set
the
HP
3457A
to
the
4-Wire
Ohms
Function
(OHMF)
and
the
number
of
digits
displayed
(DIGITS
DISP)
to
six.
2.
Short
the
front
panel
input
terminals
of
the
HP
3457A
as
shown
in
Figure
4-17.
3.
Use
the
Test
Record
to
record
the
offset
readings
of
the
3
Kohm,
30
Kohm
and
300
Kohm
ranges.
4.
If
any
of
the
offset
readings
are
beyond
the
limits
specified
in
Table
4-27
and
on
the Test
Record,
the
instrument
should
be
calibrated.
Refer
to
Section
V
of
this
manual
for
calibration
procedures.
5.
Remove
the
short
from
the
front
panel
input
terminals.
4.29