HP 3457A - 4-86. 4-Wire Ohms Function - Gain Test

HP 3457A
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operational
Verification
Tests
4-Wire
Ohms
Operational
Verification
Tests
Cont’d
34578
3457A
|3457A
|
90
day
Limits
|
1
year
Limits
Input
Range
|
Set
Up
Short
3
Kohm
|
OHMF
|
0.000007
Kohm
|
0.000007
Kohm
Short
30
Kohm
|
OHMF
|
00.00007
Kohm
|
00.00007
Kohm
Short
|300
Kohm
|
OHMF
|
000.0008
Kohm
|
000.0008
Kohm
Table
4-27.
4-Wire
Ohms
Offset
Test
Limits
Figure
4-18.
4-Wire
Ohms
Operational
Verification
Test
Connections
4-86.
4-Wire
Ohms
Function
-
Gain
Test
3457A
4-87.
Equipment
Required.
Resistance
Standards
of
3
Kohms
(+
.001%),
30
Kohms
(+
001%)
and
300
Kohms
(+
,001%)
are
required
for
this
procedure.
I.
Set
the
HP
3457A
to
the
4-Wire
ohms
function
(OHMF)
and
the
number
of
digits
displayed
(DIGITS
DISP)
to
six.
2.
Connect
the
appropriate
Resistance
Standard
to
the
HP
3457A
front
panel
input
terminals
as
shown
in
Figure
4-18.
3.
Use
the
Test
Record
to
record
the
resistance
readings
for
the
3
Kohm,
30
Kohm
and
300
Kohm
ranges.
4.
If
any
of
the
readings
are
beyond
the
limits
specified
in
Table
4-28
and
on
the
Test
Record,
the
in-
strument
should
be
calibrated.
Refer
to
Section
V
of
this
manual
for
calibration
procedures.
5.
Remove
the
Resistance
Standard
from
the
HP
3457A
front
panel
input
terminals.

Other manuals for HP 3457A