Operational
Verification
Tests
3457A
2-Wire
Ohms
Operational
Verification
Tests
Cont’d
SENSE
INPUT
TAST
a
im
fri
FILTER
CIC
Figure
4-16.
2-Wire
Ohms
Operational
Verification
Test
Connections
4.
If
any
of
the
offset
readings
are
beyond
the
limits
specified
in
Table
4-25
and
on
the
Test
Record,
the
instrument
should
be
calibrated.
Refer
to
Section
V
of
this
manual
for
calibration
procedures.
5.
Remove
the
short
from
the
front
panel
input
terminals.
34570
34570 34578
90
day
limits
1
year
limits
Input
Range
|
Set
Up
Short
3
Kohm
|
OHM
0.000207
Kohm
|
0.000207
kohm
Short
30
Kohm
|
OHM
00.00027
Kohm
|
00.00027
Kohm
Short
|300
Kohm
|
OHM
000.0010
Kohm
|
000.0010
Kohm
Table
4-25.
2-Wire
Ohms
Offset
Test
Limits
4-82.
2-Wire
Ohms
Function
-
Gain
Test
4-83.
Equipment
Required.
Resistance
Standards
of
3
Kohms
(+
.001%),
300
Kohms
(+
001%)
are
required
for
this
procedure.
30
Kohms
(+
001%),
and
LSet
the
HP
3457A
to
the
2-wire
ohms
function
(OHM)
and
the
number
of
digits
displayed
(DIGITS
DISP)
to
six.
2.
Connect
the
Resistance
Standard
to
the
HP
3457A
front
panel
HI
and
LO
input
terminals.
(The
connecting
wires
should
be
as
short
as
possible
to
reduce
lead
resistance).
3.
Use
the
Test
Record
to
record
the
3
Kohm,
30
Kohm
and
300
Kohm
full-scale
readings.
4.
If
any
of
the
readings
are
beyond
the
limits
specified
in
Table
4-26
and
on
the
Test
Record,
the
in-
strument
should
be
calibrated.
Refer
to
Section
V
of
this
manual
for
calibration
procedures.
5.
Remove
the
Resistance
standard
from
the
HP
3457A
front
panel
input
terminals.