HP 3457A - Rear Input Performance Tests (44491 A General Purpose Relay Assy); 4-50. DC Voltage Function - 44491 A Offset Test

HP 3457A
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance
Tests
3457A
Rear
Input
Performance
Tests
(44491A
General
Purpose
Relay
Assy)
4
WIREO
10
aL
iz
13
tT
EAWIRE
44491A
RELAY
SHORTING
WIRES
Figure
4-10.
HP
44491A
Terminal
Block
Assembly
Connections
4-49.
Preliminary
Steps
1.
Short
the
Rear
Input
HI
and
LO
terminals
and
the
Q
Sense
HI
and
LO
terminals
as
shown
in
Figure
4-9,
2.
Select
the
Rear
Terminal
Input.
(Press
the
TERM
configuration
key,
enter
the
number
2
and
press
the
ENT
key.
NOTE
The
44491A
Relay
Assembly
and
the
3457A
Multimeter
must
have
been
calibrated
as
a
unit
for
these
tests
to
apply.
Calibration
procedures
are
provided
in
Section
V
of
this
manual.
4-50.
DC
Voltage
Function
-
44491A
Offset
Test
4-51.
Equipment
Required.
Low
thermal
(copper)
shorting
wires
are
required
for
this
procedure.
1,
Set
the
HP
3457A
to
the
DC
Voltage
function
(DCV)
and
the
number
of
digits
displayed
(DIGITS
DISP)
to
six.
2.
Close
channel
0,
(Press
the
CHAN
configuration
key,
enter
the
number
0
and
press
the
ENT
key).
3.
Test
the
input
offset
on
the
300
V,
30
V,
3
V,
300
mV
and
30
mV
ranges
and
record
the
reading
of
each
range
on
the
Test
Record
provided
at
the
end
of
this
section.
Begin
with
the
300
V
range
to
al-
low
any
thermal
voltages
which
might
affect
the
readings
on
the
lower
ranges
to
dissipate.
4,
Test
the
HP
3457A
input
offset
on
channels
0
through
7
with
the
instrument
set
to
the
30
mV
range.
Record
the
reading
of
each
channel
on
the
Test
Record
provided
at
the
end
of
this
section.
5.
If
any
of
the
offset
readings
are
beyond
the
limits
specified
in
Table
4-15
and
on
the
Test
Record,
the
instrument
should
be
calibrated
or
repaired.
Refer
to
Section
V
for
calibration
procedures.