EasyManua.ls Logo

HP 3457A - 4-54. 4-Wire Ohms Function - 44491 A Offset Test Limits; 4-56. DC Current Function - 44491 A Offset Test Limits

HP 3457A
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3B457A
Performance
Tests
Rear
Input
Performance
Tests
(44491A
Relay
Assy.)
Cont’d
2.
Close
channel
10.
(Press
the
CHAN
configuration
key,
enter
the
number
10
and
press
the
ENT
key).
3.
Use
the
Test
Record
to
record
the
offset
readings
of
the
30
Ohm,
300
Ohm,
3
Kohm,
30
Kohm,
300
Kohm,
3
Mohm,
and
30
Mohm
ranges.
4.
If
any
of
the
offset
readings
are
beyond
the
limits
specified
in
Table
4-17
and
on
the
Test
Record,
the
instrument
should
be
calibrated
or
repaired.
Refer
to
Section
V
for
calibration
procedures,
34578 3457A
|
34578
Test
Limits
Input
Range
|
Set
Up
Short
30
ohm
OHMF
|
00.00635
ohm
Short
|300
ohm
oHMF
|
000.0065
ohm
Short
3
Kohm
|
OHMF
|
0.000010
Kohm
Short
30
Kohm
OHMF
00.00010
Kohm
Short
|300
Kohm
|
OHMF
|
000.0011
Kohm
Short
3
Mohm
|
OHMF
|
0.000017
Mohm
Short
30
Mohm
|
OHMF
|
00.00086
Mohn
Table
4-17.
4-Wire
Ohms
Offset
Test
Limits
(44491A)
4-56.
DC
Current
Function
-
44491A
Offset
Test
Limits
4-57.
Equipment
Required.
This
procedure
does
not
require
any
test
equipment.
1.
Set
the
HP
3457A
to
the
DC
Current
function
(DCI),
300
uA
range,
and
the
number
of
digits
dis-
played
(DIGITS
DISP)
to
six.
2.
Use
the Test
Record
to
record
the
current
offset
readings
on
the
300 uA,
3
mA,
30
mA,
300
mA
and
|
A
ranges.
3.
If
any
of
the
offset
readings
are
beyond
the
limits
specified
in
Table
4-18
and
on
the
Test
Record,
the
instrument
should
be
calibrated
or
repaired.
Refer
to
Section
V
for
calibration
procedures.
3457A
|3457A
|
3457A
Test
Limits
Input
|Range
|Set
Up
High
Low
Open
|300
ua]
ocr
+000.0104
uA
-000.0104
uA
Open
3
mA}
DCI
+0.000104
mA
-0.000104
mA
open
|
30
ma|
oci
+00.00104
mA
-00.00104
ma
Open
|300
mA]
DCI
+000.0204
mA
-000.0204
mA
Open
1A
|
pcr
+0.000604
A
0.000604
A
Table
4-18.
DC
Current
Offset
Test
Limits
(44491A)
4-19