Performance
Tests
3457A
DC
Current
Performance
Tests
Cont'd
3457A
|3457A
|
3457A
90
day
limits
1
year
limits
Input
Range
|Set
Up
High
Low
High
Low
Open
|
300
uA]
DCI
+000.0104
uA
-000.0104
uA
+000.0104
uA
-000.0104
uA
Open
3
mA}
DCI
+0.000104
mA
-0.000104
mA
+0.000104
ma
-0.000104
mA
Open
30
mA}
DCI
+00.00104
mA
-00.00104
ma
+00.00104
mA
-00.00104
mA
Open
|300
mA]
OCI
+000.0204
ma
-000.0204
ma
+000.0204
mA
-000.0204
ma
Open
1A
ocl
+0.000604
A
-0.000604
A
+0.000604
A
-0.000604
A
300
uA}]300
uA}
DCI
300.0704
uA
299.9296
uA
300.1304
uA
299.8696
uA
3
mA
3
mA}
DCI
3.000704
ma
2.999296
mA
3.001304
mA
2.998696
mA
30
mA|
30
mA|
DCI
30.00704
mA
29.99296
mA
30.01304
mA
29.98696
mA
300
mA|300
mA}
DCI
300.2304
mA
299.7696
mA
300.2604
mA
299.7396
mA
1A
1A
ocl
1.001304
A
0.998696
A
1.001304
A
0.998696
A
Table
4-5.
DC
Current
Offset
and
Gain
Test
Limits
4-24,
DC
Current
Function
-
Offset
Test
4-25.
Equipment
Required.
This
procedure
does
not
require
any
test
equipment.
1.
Set
the
HP
3457A
to
the
DC
Current
function
(DCI),
300
uA
range,
and
the
number
of
digits
dis-
played
(DIGITS
DISP)
to
six.
2.
Use
the
Test
Record
to
record
the
current
offset
readings
on
the
300 uA,
3
mA,
30
mA,
300
mA
and
|
A
ranges.
3.
If
any
of
the
offset
readings
are
beyond
the
limits
specified
in
Table
4-5
and
on
the
Test
Record,
the
instrument
should
be
calibrated
or
repaired.
Refer
to
Section
V
of
this
manual
for
calibration
procedures.
4-26.
DC
Current
Function
-
Gain
Test
4-27.
Equipment
Required.
A
DC
Current
Source
capable
of
providing
currents
of
300
uA,
3
mA,
30
mA
(+
007%),
300
mA
(+
.026%),
and
1
A
(+
.04%)
is
required
for the
following
procedure.
4-8
1.
Set
the
DC
Current
Source
to
300
uA
and
connect
it
to
the
HP
3457A
front
panel
I
and
LO
input
terminals.
2.
Set
the
HP
3457A
to
the
300
uA
range
and
the
number
of
digits
displayed
(DIGITS
DISP)
to
six.
3.
Use
the
Test
Record
to
record
the
full-scale
current
readings
for
the
300 uA,
3
mA,
30
mA,
300
mA
and
|
A
ranges.
4.
If
any
of
the
full-scale
readings
are
beyond
the
limits
specified
in
Table
4-5
and
on
the
Test
Record,
the
instrument
should
be
calibrated
or
repaired.
Refer
to
Section
V
of
this
manual
for
calibration
procedures.
5.
Reduce
the
output
of
the
DC
Current
Source
and
disconnect
it
from
the
HP
3457A
input
terminals.