HP 3457A - AC Current Performance Tests; 4-28. AC Current Function - Gain Test

HP 3457A
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AC
Current
Performance
Tests
Performance
Tests
SENSE
INPUT
(a
Mine:
(2
MIRED
Cc
|©@-©)|
ERROR
ens,
HATH
HATH
RECALL
STORE
ALMEAN R_SDEV
R_NSAMP
AC
CURRENT
SOURCE
Figure
4-4.
ACI
Performance
Test
Connections
4-28.
AC
Current
Function
-
Gain
Test
3.
Use
the
Test
Record
to
record
the
30
mA,
300
mA
and
|
A
full-scale
readings.
34578
|
3457A
|
34578
90
day
Limits
Input
|Range
|
Set
Up
High
|
Low
30
mA|
30
mA]
ACI
30.10300
mA
29.89700
mA
300
ma|300
ma]
ACI
301.0300
ma
298.9700
ma
ta]
1A]
act
1.005300
A
0.994700
A
Table
4-6.
AC
Current
Test
Limits
4-29.
Equipment
Required,
An
AC
Current
Source
capable
of
providing
currents
of 30
mA
(+
14%),
300
mA
(+
.14%)
and
1
A
(+
.24%)
at
a
frequency
between
100
Hz
and
20
KHz
is
required
for
the
follow-
1.
Set
the
AC
Current
source
for
an
output
of
30
mA
at
a
frequency
between
100
Hz
and
20
KHz
and
connect
it
to
the
HP
3457A
front
panel
I
and
LO
input
terminals.
2.
Set
the
HP
3457A
to
the
30
mA
range
and
the
number
of
digits
displayed
(DIGITS
DISP)
to
six.
4.
If
the
full-scale
readings
are
beyond
the
limits
specified
in
Table
4-6
and
on
the
Test
Record,
the
in-
strument
should
be
calibrated
or
repaired.
Refer
to
Section
V
of
this
manual
for
calibration
5.
Reduce
the
output
of
the
AC
Current
Source
and
disconnect
it
from
the
HP
3457A
input
terminals.