8-84.
44492A
If
the
instrument
does
not
work
when
the
card
is
installed:
a)
Check
the
+5
volt
supply
coming
into
the
board
at
connector
P851.
b)
Cut
jumpers
JM851/JM852.
If
the
instrument
begins
working
replace
U852.
If
it
does
not
begin
working
replace
U85I.
These
shift
registers
can
load
down
the
serial
data
lines
SDS,
SDO,
or
SDC2
which
will
prevent
the
instrument
from
working.
c)
Replace
jumpers
JM851/JM852.
If
the
instrument
works
but the
channels
are
not
operating
correctly,
the
following
items
should
be
checked:
1)
Check
for
the
correct
output
from
U851
and
U852.
All
outputs
should
be
high
except
for
the
output
to
the
relay
being
closed.
To
close
relay
K851
(channel
0)
requires
that
output
QI
of
shift
register
U851
be
held
low.
The
drive
signal
must
remain
low
the
entire
time
the
relay
is
closed.
2)
If
the
output
of
the
shift
registers
is
correct,
check
the
output
of
drive
transistors
U853-U855,
If
relay
851
is
closed,
the
output
of
U854
pin
14
should
be
approximately
+5
volts.
If
it
is
not,
replace
the
transistor
pack.
3)
If
the
drive
signal
is
working
and
the
contacts
do
not
operate
correctly,
replace
the
relay.
Contact
operation
can
be
checked
by
making
a
resistance
measurement
from
the
correct
pin
of
connector
P853
to
the
correct
pin
of
connector
P852.
To
check
the
high
contact
of
relay
K851,
resistance
should
be
measured
between
connector
P853
pin
|
and
connector
P852
pin
22.
8-85.
Replacement
Procedures
Procedures
for the
removal
and
replacement
of
the
PC
assemblies
are
contained
in
Section
VI
of
this
manual.
Please
observe
clean
handling
techniques
and
anti-static
procedures
when
handling
the
PC
assemblies.
3457A