SECTION
III
OPERATION
3-1.
INTRODUCTION
3-2.
The
information
contained
in
this
section
describes
manual
operation
of
the
HP
3457A
Multimeter,
The
contents
are
intended
to
meet
the
needs
of
service
personnel
with
emphasis
on
front
panel
operation,
For
more
detailed
information
on
overall
operation,
refer
to
the
HP
3457A
Operating
Manual.
3-3.
GENERAL
INFORMATION
3-4.
AC
Power
3-5.
Before
connecting
power
to
the
HP
3457A,
be
certain
the
line
switches
on
the
rear
panel
are
set
for
your
power
source
and
that
the
proper
fuse
is
installed.
Information
for
setting
the
line
switches
and
for
fuse
selection
is
located
in
Section
II
of
this
manual.
3-6.
Power-On
3-7.
The
HP
3457A
automatically
performs
a
power-on
self
test
when
it
is
switched
on,
This
test
takes
ap-
proximately
1.5
seconds
to
complete
and
basically
assures
that
the
instrument
is
capable
of
operating.
The
test
does
not
necessarily
indicate
that
measurements
will
be
accurate.
The
power-on
self
test
checks
the
master
processor,
slave
processor
and
the
communication
isolation
circuitry
(three
of
the
ten
tests
perform-
ed
by
the
complete
self
test).
Upon
satisfactory
completion
of
the
test,
the
instrument
will
display
its
HP-IB
address
for
approximately
five
seconds.
The
address
is
displayed
as
ADDRESS-dd
where
"dd"
is
the
decimal
address
code
of
the
instrument
(the
address
is
set
to
decimal
22
at
the
factory).
If
the
test
can
not
be
com-
pleted,
the
instrument
will
display
FAILED.
If
this
occurs,
refer
to
Section
VIII
for
troubleshooting
procedures,
3-8.
Upon
completion
of
the
Power-On
Test,
the
instrument
sets
itself
to
predefined
conditions
(Power-On
state).
The
power-on
state
conditions
are
listed
in
Table
3-1.
3-9.
OPERATION
3-10.
The
following
paragraphs
des
sic
Operating
procedures
for the
HP
3457A
Multimeter,
This
sec-
tion
covers
only
front
panel
operating
procedures
such
as
voltage,
resistance,
and
current
measurements.
Also
included
are
manual
procedures
covering
operation
of
the
HP
44491A
and
44492A
Multiplexer
Assemblies.
For
detailed
operating
information,
refer
to
the
HP
3457A
Operating
Manual.
3-11.
Voltage
Measurements
(Manual
Operation)
3-12.
Function
Selection.
The
measurement
function
is
selected
by
pressing
the
appropriate
key.
The
volt-
age
measurement
functions
available
are
DC
Volts,
AC
Volts
and
DC
Coupled
AC
Volts.
The
default
func-
tion
is
DC
Volts.
®
To
measure
DC
Voltage
-
Press
the
DCV
function
key.
©
To
measure
AC
Voltage
-
Press
the
ACV
/
ACDCV
function
key.
©
To
measure
DC
coupled
AC
voltage
-
Press
the
Blue
SHIFT
key
then
the
ACV
/
ACDCV
function
key.
3-1