HP 3457A - Rear Input Performance Tests (Standard Instrument); 4-91. DC Voltage Function - Rear Terminal Offset Test

HP 3457A
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operational
Verification
Tests
Rear
Input
Operational
Verification
Tests
(Standard
Instrument)
REAR
TERMINALS
450V
pk
MAX
TO
FRONT
TERMINALS
ORS
CUSENSE
(4
wiRE)
‘INPUT
(2
WIRE)
HI
LO
HI
LO
AX
Figure
4-20.
Standard
Instrument
Rear
Input
Test
Connections
4-90.
Preliminary
Steps
1.
Short
the
Rear
Input
HI
and
LO
terminals
and
the
Q
Sense
HI
and
LO
terminals
as
shown
in
Figure
4-9,
2.
Select
the
Rear
Terminal
Input.
(Press
the
TERM
configuration
key,
enter
the
number
2
and
press
the
ENT
key
4-91.
DC
Voltage
Function
-
Rear
Terminal
Offset
Test
4-92.
Equipment
Required.
A
low
thermal
short
(copper
wire)
is
required
for
this
procedure,
lL.
Set
the
HP
3457A
to
the
DC
Voltage
function
(DCV)
and
the
number
of
digits
displayed
(DIGITS
DISP)
to
six.
2.
Test
the
HP
3457A
input
offset
on
the
30
V,
3
V
and
300
mV
ranges
and
record
the
reading
of
each
range
on
the
Test
Record
provided
at
the
end
of
this
section.
Begin
with
the
30
V
range
to
allow
any
thermal
voltages
which
might
affect
the
readings
on
the
lower
ranges
to
dissipate.
3.
If
any
of
the
offset
readings
are
beyond
the
limits
specified
in
Table
4-29
and
on
the
Test
Record,
the
instrument
should
be
calibrated.
Refer
to
Section
V
of
this
manual
for
calibration
procedures.
3457A
3457A
134570
|
34570
90
day
limits
1
year
Limits
Input
|Range
|
Set
Up
High
Low
High
Low
Short
|
30
v
|
pcv
+
00.00020
v
-
00.00020
v
+
00.00020
v
-
00.00020
v
Short
3v
|
pev
+
0.000007
v
0.000007
v
+
0.000007
v
0.000007
v
Short
|300
mv|
pcv
+000.0040
mv
-000.0040
mv
+000.0040
mv
=000.0040
mv
4-32
Table
4-29.
DC
Voltage
Offset
Test
Limits
(Rear
Terminals)