3457A
Calibration
e.
Repeat
steps
"dil"
through
"
d4"
for
each
of
the
remaining
ranges.
&
5-59.
Two-Wire
Ohms
Offset
Calibration
-
HP
44491A
a,
Select
the
2-Wire
Ohms
function.
(Press
the
OHM
function
key
).
b.
Close
channel
0.
(Press
the
CHAN
configuration
key,
enter
the
number
0
and
press
the
ENT
key
).
c.
Use
the
following
procedure
to
calibrate
the
2-wire
ohm
offset
on
the
30
Mohm,
3
Mohm,
300
Kohm,
30
Kohm,
3
Kohm,
300
Ohm
and
30
Ohm
ranges
beginning
with
the
30
Mohm
range.
I,
Set
the
3457A
to
the
appropriate
resistance
range.
(Use
the
{
or
|
scroll
keys
to
select
the
proper
range),
2.
Access
the
Calibration
Command.
(Press
the
Blue
SHIFT
key,
then
the
OFFSET
COMP
/
C
configura-
tion
key).
The
display
will
show
CAL,
3.
Enter
the
value
of
the
input
-
in
this
case
0.
(NOTE:
If
the
security
feature
has
not
been
disabled
(set
to
0),
it
will
be
necessary
to
enter
the
input
value,
the
delimiter
(,
)
and
the
security
code).
4.
Press
the
ENT
key.
The
display
will
show
CALIBRATING
while
the
calibration
routine
is
running.
d,
Repeat
steps
"cl
"
through
"c4
"
for
each
of
the
remaining
ranges.
5-60.
Four-Wire
Ohms
Offset
Calibration
-
HP
44491A
e
a.
Select
the
4-Wire
Ohms
function.
(Press
the
Blue
SHIFT
key
then
the
OHM
/
OH
MF
function
key
).
b.
Close
channel
10.
(Press
the
CHAN
configuration
key,
enter
the
number
10
and
press
the
ENT
key
).
c.
Use
the
following
procedure
to
calibrate
the
4-Wire
ohm
offset
on
the
30
Mohm,
3
Mohm,
300
Kohm,
30
Kohm,
3
Kohm,
300
Ohm
and
30
Ohm
ranges
beginning
with
the
30
Mohm
range.
1.
Set
the
3457A
to
the
appropriate
resistance
range.
(Use
the
t
or
J
scroll
keys
to
select
the
proper
range).
2.
Access
the
Calibration
Command.
(Press
the
Blue
SHIFT
key,
then
the
OFFSET
COMP
/
C
configura-
tion
key).
The
display
will
show
CAL.
3.
Enter
the
value
of
the
input
-
in
this
case
0,
(NOTE:
If
the
security
feature
has
not
been
disabled
(set
to
0),
it
will
be
necessary
to
enter
the
input
value,
the
delimiter
(
,
)
and
the
security
code).
4.
Press
the
ENT
key.
The
display
will
show
CALIBRATING
while
the
calibration
routine
is
running.
d.
Repeat
steps
"cl
"
through
"c4
"for
each
of
the
remaining
ranges.
5-61.
DC
Current
Offset
Calibration
-
HP
44491A
a.
Select
the
DC
Current
function.
(Press
the
DCI
function
key.)
&
b.
Select
the
Scanner
Input.
(Press
the
TERM
configuration
key,
enter
the
number
2,
and
press
the
ENT
key.)