HP 3457A - 5-37. DC Current Gain Calibration - Front Terminals

HP 3457A
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
3457A
d.
Repeat
steps
"cl
"
through
"c5"
for
each
of
the
remaining
ranges.
e,
Disconnect
the
resistance
standard
from
the
3457A
input
terminals.
5-35.
DC
Current
Offset
Calibration
-
Front
Terminals
5-36.
Equipment
Required.
No
equipment
is
needed
for
this
procedure.
a.
Be
certain
all
leads
are
disconnected
from
the
HP
3457A
input
terminals.
b.
Select
the
DC
Current
Function.
(Press
the
DCI
function
key).
c.
Use
the
following
procedure
to
calibrate
the
front
terminal
offset
on
the
1
A,
300
mA,
30
mA,
3
mA
and
300
uA
ranges
beginning
with
the
1
A
range.
1.
Set the
3457A
to
the
appropriate
current
range.
(Use
the
it
or
i
scroll
key
to
select
the
proper
range),
2.
Access
the
Calibration
Command.
(Press
the
Blue
SHIFT
key,
then
the
OFFSET
COMP
/
C
configura-
tion
key).
The
display
will
show
CAL.
3.
Enter
the
value
of
the
input
-
in
this
case
0.
(NOTE:
If
the
security
feature
has
not
been
disabled
(set
to
0),
it
will
be
necessary
to
enter
the
Input
value,
the
delimiter
(
,
)
and
the
security
code).
4,
Press
the
ENT
key.
The
display
will
show
CALIBRATING
while
the
calibration
routine
is
running.
d.
Repeat
steps
"cl
"
through
"c4"
for
each
of
the
remaining
ranges.
5-37.
DC
Current
Gain
Calibration
-
Front
Terminals
5-38.
Equipment
Required.
A
DC
Current
Source
capable
of
providing
300
uA,
3
mA,
30
mA
(+.007%),
300
mA
(+,026%)
and
|
A
(+,04%)
is
required
to
calibrate
the
DC
Current
Function.
a.
Connect
the
Current
Source
to
the
HP
3457A
front
panel
I
and
LO
input
terminals.
Set
the
output
of
the
Current
Source
to
|
A.
b.
Use
the
following
procedure
to
calibrate
the
front
terminal
gain
on
the
|
A,
300
mA,
30
mA,
3
mA
and
300
uA
ranges
beginning
with
the
|
A
range.
1,
Set
the
3457A
to
the
appropriate
current
range.
(Use
the
{
or
|
scroll
key
to
select
the
proper
range).
2.
Set
the
current
source
to
provide
a
full-scale
reading
on
the
HP
3457A
range
selected.
3.
Access
the
Calibration
Command.
(Press
the
Blue
SHIFT
key,
then
the
OFFSET
COMP
/
C
con-
figuration
key
).
The
display
will
show
CAL.
4.
Enter
the
value
of
the
input
current
(in
amps).
(NOTE:
If
the
security
feature
has
not
been
disabled
(set
to
0),
it
will
be
necessary
to
enter
the
input
value,
the
delimiter
(
,
)
and
the
security
code).
5.
Press
the
ENT
key.
The
display
will
show
CALIBRATING
while
the
calibration
routine
is
running.
c.
Repeat
steps
"
bl
"
through
"
b4
"
for
each
of
the
remaining
ranges.
5-10