HP 3457A - AC Voltage Operational Verification Tests; 4-70. AC Voltage Function - Gain Test

HP 3457A
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operational
Verification
Tests
AC
Voltage
Operational
Verification
Tests
Figure
4-13.
ACV
Operational
Verification
Test
Connections
4-70.
AC
Voltage
Function
-
Gain Test
3457A
4-71,
Equipment
Required.
An
AC
Signal
Source
capable
of
providing
sine-wave
voltages
of
300
mV,
3
V,
and
30
V
(+
0.05%)
at a
frequency
of
|
KHz
is
required
for
the
following
procedure.
1,
Set
the
HP
3457A
to
the
AC
Voltage
function
(ACV).
2.
Connect
the
AC
Signal
Source
to
the
HP
3457A
front
panel
HI
and
LO
input
terminals.
Set
the
Signal
Source
for
an
output
voltage
of 30
V
at a
frequency
of
1
KHz.
3.
Use
the
Test
Record
to
record
the
300
mV,
3
V
and
30
V
full-scale
readings
beginning
with
the
30
V
range.
4. If
any
of
the
gain
reading
are
beyond
the
limits
specifies
3B457A
Input
|3457a
|
34570
90
day
limits
Input Freq.
|Range
|
Set
Up
High
Low
300
mv
1
kHz
[300
mv]
acy
300.5020
mv
299.4980
mv
3V
1
kHz
3.v
|]
acy
3.005100
v
2.994900
V
30
Vv
1
kuz
|
30
Vv
|
acy
30.05020
v
29.94980
V
Table
4-22.
AC
Gain Test
Limits
4-72.
AC
Voltage
Function
-
Frequency
Response
Test
1
in
Table
4-22
and
on
the
Test
Record,
the
instrument
should
be
calibrated
or
repaired.
Refer
to
Section
V
for
calibration
procedures.
4-73.
Equipment
Required.
An
AC
Signal
Source
capable
of
providing
sine-wave
voltages
of
300
mV
and
3
V
(£.2%)
at
20
KHz,
300
mV
and
3
V
(+.1%)
at
100
Hz
and
300
mV
and
3
V
(4.3%)
at
20
Hz
is
required
for
this
test.
|.
Connect
the
AC
Signal
Source
to
the
HP
3457A
front
panel
HI
and
LO
input
terminals.
Set
the
AC
Voltage
Standard
for
an
output
voltage
of
300
mV
at
a
Frequency
of
20
KHz.