HP 3457A - 5-43. Frequency Calibration - Front Terminals; 5-45. CALIBRATION PROCEDURES - REAR PANEL INPUTS; 5-47. DC Volts Offset Calibration - Rear Terminals (Standard Instrument)

HP 3457A
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
3457A
c.
Use
the
following
procedure
to
calibrate
the
AC
Current
function
on
the
30
mA,
300
mA
and
|
A
ranges
beginning
with
the
30
mA
range.
S
1,
Set
the
3457A
to
the
appropriate
current
range.
(Use
the
{
or
|
scroll
key
to
select
the
proper
range).
2.
Set
the
current
source
to
provide
a
full-scale
reading
on
the
HP
3457A
range
selected.
3.
Access
the
Calibration
Command.
(Press
the
Blue
SHIFT
key,
then
the
OFFSET
COMP
/
C
configura-
tion
key).
The
display
will
show
CAL.
4.
Enter
the
value
of
the
input
current
(in
amps).
(NOTE:
If
the
security
feature
has
not
been
disabled
(set
to
0),
it
will
be
necessary
to
enter
the
input
value,
the
delimiter
(
,
)
and
the
security
code).
5,
Press
the
ENT
key.
The
display
will
show
CALIBRATING
while
the
calibration
routine
is
running.
d,
Repeat
steps
"cl"
through
"c5"
for
each
of
the
remaining
ranges.
e.
Set
the
Current
Source
for
an
output
of
0
and
disconnect
it
from
the
HP
3457A,
5-43.
Frequency
Calibration
-
Front
Terminals
5-44.
Equipment
Required.
A
Frequency
Generator
capable
of
providing
a
single
sine-wave
signal
between
100
Hz
and
|
MHz
with
a
frequency
accuracy
of
+.003%)
is
required
to
calibrate
the
Frequency
Function.
a,
Set
the
3457A
to
the
Frequency
function,
(Press
the
FREQ
function
key
).
b,
Set
the
Frequency
Source
to
a
known
frequency
between
100
Hz
and
|
MHz
and
connect
it
to
the
a
HP
3457A
front
panel
HI
and
LO
input
terminals.
c.
Access
the
Calibration
command.
(Press
the
Blue
SHIFT
key,
then
the
OFFSET
COMP
/
C
configuration
key).
The
display
will
show
CAL.
d.
Enter
the
value
of
the
input
frequency
(in
hertz).
(NOTE:
If
the
security
feature
has
not
been
disabled
(set
to
0),
it
will
be
necessary
to
enter
the
input
frequency,
the
delimiter
(
,
)
and
the
security
code),
e.
Press
the
ENT
key.
The
display
will
show
CALIBRATING
while
the
calibration
routine
is
running.
f.
Disconnect
the
Frequency
Source
from
the
HP
3457A,
5-45.
CALIBRATION
PROCEDURES
-
REAR
PANEL
INPUTS
5-46.
The
following
paragraphs
describe
the
calibration
process
for the
HP
3457A
rear
panel
inputs.
Separate
procedures
are
provided
for
the
standard
instrument
(Rear
Input
Terminals),
and
instruments
equipped
with
optional
plug-in
assemblies
44491A
(General
Purpose
Relay
Assembly)
and
44492A
(Reed
Relay
Multiplex
Assembly).
Select
the
procedure
which
applies
to
your
instrument.
5-47.
DC
Volts
Offset
Calibration
-
Rear
Terminals
(Standard
Instrument)
5-48.
Equipment
Required.
A
low
thermal
short
(copper
wire)
is
required
for
this
procedure.
a.
Short
the
Q
SENSE
HI
and
LO
and
INPUT
HI
and
LO
rear
input
terminals
together.
5-12