pdhman.book : pdhapp.doc 53 Tue Oct 14 15:00:20 1997
53
PDH / DSn Testing With The HP 37717C
Frequency Offset Tolerance
2. Set up the display as shown
opposite.
Select the PATTERN required from the
menu.
3. Set up the display as shown
opposite.
Any of the other results can be selected
without affecting the measurement.
Continuity Check
Before running the test carry out a continuity test to verify the measurement path.
1. Connect a loopback at the desired point on the line terminal or cross-multiplexer
equipment.
2. Press to start a measurement.
3. Press error add three times and check that the errors are recorded on the
display.
4. Press to stop the measurement.
RECEIVE
RESULTS
RUN/STOP
SINGLE
RESULTS
RUN/STOP