Figure 2-5. Power Sweep Linearity Test Setup
4.
Press
4
Meas
5
,
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ANALYZER TYPE
,
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NETWORK ANALYZER
,
4
Preset
5
to initialize the HP 4395A. Then
set the controls as follows:
Control Settings Key Strokes
Power Sweep
4
Sweep
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SWEEP TYPE MENU
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POWER SWEEP
CW Frequency: 50 MHz
4
Source
5
,
NNNNNNNNNNNNNNNNNNNNNNN
CW FREQ
,
4
5
5
,
4
0
5
,
4
M/
5
Start Power:
0
50 dBm
4
Start
5
,
4
-
5
,
4
5
5
,
4
0
5
,
4
x1
5
Stop Power:
0
30 dBm
4
Stop
5
,
4
-
5
,
4
3
5
,
4
0
5
,
4
x1
5
Number of Points: 5
4
Sweep
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NUMBER of POINTS
,
4
5
5
,
4
x1
5
Trigger: Manual
4
Trigger
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRIGGER[xxx]
,
NNNNNNNNNNNNNNNNNNNN
MANUAL
Trigger Event:On Point
4
Trigger
5
,
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRIGGER[xxx]
,
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRIG EVENT [ON SWEEP]
(Then the softkey label changes to
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRIGGER [ON POINT]
5.
Press
4
Trigger
5
,
NNNNNNNNNNNNNNNNNNNN
SINGLE
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRIGGER:[MANUAL]
,
NNNNNNNNNNNNNNNNNNNN
MANUAL
to start a power sweep and to set the
HP 4395A power to the rst sweep point of
0
50 dBm listed in Table 2-4. Table 2-4 lists test
settings.Verify that the step attenuator is set to 10 dB
.
Table 2-4. Power Sweep Linearity Test Settings 1
HP 4395A
Source Power
Step Attenuator
0
50 dBm 10 dB
0
45 dBm 10 dB
0
40 dBm 10 dB
0
35 dBm 10 dB
0
30 dBm 10 dB
6. Wait for the power meter reading to settle
.
2-12 Performance Tests