19. Frequency and Sensitivity Check - 500 MHz to 18 GHz 
 a. Performance Check 
CAUTION 
Do not exceed +25 dBm peak of input power at the N type 
connector (500 MHz to 18 GHz), SMA connector for model 
5343A.   Damage to the internal sampler may result. 
NOTE 
When the input signal level to the high frequency input 
connector exceeds approximately +5 dBm, each digit in the 
display becomes a minus sign (-) to indicate overload. 
 (1) Connect signal generator RF Output to TI N type input.  (Use APC 3.5 male 
connector with collar for model 5343A). 
 (2) Set 10 Hz - 500 MHz/500 MHz - 18 GHz (10 Hz - 500 MHz/500 MHz - 26.5 
GHz for model 5343A) switch to 500 MHz - 18 GHz (500 MHz - 26.5 GHz for model 
5343A). 
 (3) Press FREQ SHIFT (blue) and RESOLUTION 1 Hz pushbuttons. 
 (4) Set signal generator frequency to 500 MHz and level output to -25 dBm (-33 dBm for 
model 5343A).  TI will display a stable indication of applied frequency. 
 (5) Adjust signal generator frequency from 500 MHz to 12.4 GHz.  TI will display a 
stable indication of applied frequency. 
 (6) Set signal generator frequency to 12.4  GHz and level output to -20 dBm (-28 dBm for 
model 5343A).  TI will display a stable indication of the applied frequency. 
 (7) Adjust signal generator frequency from 12.4 to 18 GHz.  TI will display a stable 
indication of applied frequency. 
 b. Adjustments.  No adjustments can be made. 
20. Power Supply 
NOTE 
Do not perform power supply check if all other  parameters are 
within tolerance. 
 a. Performance Check.  Connect multimeter HI to -5.2 V TEST POINT (fig. 3) and 
LO to chassis ground.  If multimeter does not indicate between -5.1 and -5.25 V dc, perform 
b below.