EasyManua.ls Logo

HP 5340A - Page 9

HP 5340A
17 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TB 9-6625-2021-35
9
(7) Set signal generator frequency to 10 MHz and level output to -30 dBm. If TI
does not display a stable indication of applied frequency, perform b below.
(8) Adjust signal generator frequency from 10 to 500 MHz. If TI does not display a
stable indication of applied frequency, perform b below.
(9) Set signal generator frequency to 500 MHz and level output to -35 dBm. If TI
does not display a stable indication of applied frequency, perform b below.
(10) Adjust signal generator frequency from 500 MHz to 10 GHz. If TI does not
display a stable indication of applied frequency, perform b below.
(11) Set signal generator frequency to 10 GHz and level output to -25 dBm. If TI
does not display a stable indication of applied frequency, perform b below.
(12) Adjust signal generator frequency from 10 to 18 GHz. If TI does display a stable
indication of applied frequency, perform b below.
(13) Set RANGE switch to 250 MHz - 18 GHz.
(14) Set signal generator frequency to 250 MHz and level output to -30 dBm. TI will
display a stable indication of applied frequency.
(15) Adjust signal generator frequency from 250 to 500 MHz. TI will display a stable
indication of applied frequency.
(16) Set signal generator frequency to 500 MHz and level output to -35 dBm. TI will
display a stable indication of applied frequency.
(17) Adjust signal generator frequency from 500 MHz to 10 GHz. TI will display a
stable indication of applied frequency.
(18) Set signal generator frequency to 10 GHz and level output to -25 dBm. TI will
display a stable indication of applied frequency.
(19) Adjust signal generator frequency from 10 GHz to 18 GHz. TI will display a
stable indication of applied frequency.
b. Adjustments
(1) Connect signal generator RF Output to TI INPUT 50.
(2) Set signal generator frequency to 100 MHz and level output to -25 dBm.
(3) Adjust A17A1R22 (fig. 1) fully ccw.
(4) Adjust A17A1R11 (fig. 1) for a stable indication on TI.
(5) Decrease signal generator level output and adjust A17A1R11 (fig. 1) until
maximum sensitivity is obtained (R).
(6) Increase signal generator level output for a stable indication. Record signal
generator level output indication.
(7) Adjust A17A1R22 (fig. 1) fully cw. TI will display all zeros.

Other manuals for HP 5340A

Related product manuals