TB 9-6625-2021-35 
10 
 (8) Increase signal generator level output by 1 dB from level recorded in (6) above 
and adjust A17A1R22 (fig. 1) ccw until TI display indicates 100 MHz (R). 
 (9) Verify that sensitivity is at least -32 dBm.  
12. Power Supply 
NOTE 
Do not perform power supply check if all other parameters are 
within tolerance. 
 a. Performance Check.  Connect multimeter between test points listed in table 4 
and chassis ground.  If multimeter does not indicate within limits specified, perform 
corresponding adjustments in table 4. 
 
Table 4.  Power Supply Voltages 
 
Test instrument test 
points 
Multimeter indications  
(V) 
 
Adjustments (R) 
(fig. 1)  Min  Max  (fig. 1) 
A28TP2  −14.95  −15.05  A28R7 
A29TP2  +14.95  +15.05  A29R5 
A31TP2  −4.95  −5.05  A31R9 
A31TP3  −4.95  −5.05  A31R2 
A32TP1  +4.95  +5.05  A32R6 
A32TP2  +4.95  +5.05  A32R3 
A33TP1
1
  +10.95  +11.05  A33R8 
 
1
Option 001 only. 
 b. Adjustments.  No further adjustments can be made. 
13. Final Procedure 
 a. Deenergize and disconnect all equipment. 
 b. Annotate and affix DA label/form in accordance with TB 750-25.