Math/FFT
Menu
Vertical
Scaling
Units
Vertical
Scaling
Units
The
fundamental
measuring
units
of
an
oscilloscope
are
volts/division
on
the
vertical
axis
and
time/division
on
the
horizontal
axis.
This
philosophy
is
used
for
all
the
function
operations
except
for
FFTs.
No
provisions
were
made
to
manage
units
for
all
combinations
of
operands
and
operations.
For
example,
if
you
applied
a
+2
V
level
to
channel
1
and
a
-3
V
level
to
channel
2.
The
oscilloscope
displays
the
product
as
—6
V,
when
in
reality
it
is
-6V",
Making
FFT
Measurements
The
following
information
will
help
you
to
make
optimum
measurements
in
the
frequency
spectrum.
Connect
the
Dots
It
is
easier
to
view
FFTs
with
connect
the
dots
turned
on.
You
can
turn
connect
the
dots
on
in
the
Display
menu.
Amplitude
Measurements
For
best
vertical
accuracy
on
peak
measurements:
e@
Make
sure
the
source
impedance
and
probe
attenuation
is
set
correctly.
If
the
operand
is
a
channel,
the
impedance
and
probe
attenuation
are
set
in
the
channel
menu.
e
Set
the
source
sensitivity
so
that
the
input
signal
is
near
full
screen,
but
not
clipped.
e
Use
the
flattop
window.
e
Set
the
FFT
sensitivity
to
a
sensitive
range,
like
2
dB/division,
in
the
single-screen
mode
or
4
dB/division
in
the
dual-screen
mode.
10-15