Figure
13-2
Measurements
Automatic
Parametric
Measurements
Automatic
Parametric
Measurements
The
illustration
below shows
the
basic
process
used
by
the
oscilloscope
when
making
automatic
parametric
measurements,
like
rise
time,
Vp-p,
or
frequency.
These
measurements
can
be
made
on
input
signals,
stored
waveforms,
or
functions.
In
order
to
start
the
measurement
process,
the
oscilloscope
captures
a
data
record.
From
this
data
record,
the
oscilloscope
builds
a
histogram,
recording
how
many
times
each
q
level
(a
q
level
is
an
internal
voltage
representation)
is
present
in
the
data
record.
From
the
histogram,
and
from
the
data
record,
the
absolute
maximum
and
minimum
voltage
levels,
as
well
as
relative
maximum
(top)
and
minimum
(base)
voltage
levels,
are
determined.
Using
the
top
and
base
levels,
threshold
levels
are
calculated.
The
data
record
is
again
analyzed
using
the
thresholds
to
determine
signal
edges.
Finally,
with
all
this
information,
the
requested
parametric
measurements
are
calculated
as
shown
in
figure
13-2.
The
oscilloscope
uses
this
process
for
waveform
measurement
13-3