HP 54600 Series - Page 304

HP 54600 Series
387 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1
Using the 50-
feedthrough and the
BNC
cable, connect the signal
generator to channel 1 of the oscilloscope.
2
Set the frequency of the signal generator to 250 kHz and the
amplitude to about 800 mV.
3
Press
Autoscale
.
4
Adjust the output of the signal generator for exactly 8 divisions of
vertical deflection.
5
Change the frequency of the signal generator to the value shown
below for your instrument.
Table 4-6 Signal Generator Frequency Setting
Selected
Channel
HP 54600B HP 54601B HP 54602B * HP 54603B
Channel 1 100 MHz 100 MHz 150 MHz 60 MHz
Channel 2 100 MHz 100 MHz 150 MHz 60 MHz
Channel 3 100 MHz 250 MHz
Channel 4 100 MHz 250 MHz
*1mv/div to 5 mv/div dc = 100 MHz.
6
Change the sweep speed of the oscilloscope to 5 ns/div and observe
the display.
The vertical amplitude of the signal on the display should be equal to or
greater than 5.66 divisions (
3 dB point). If the result is not
≤−
3 dB, see
"Troubleshooting the Oscilloscope," on page 4–40.
7
Repeat steps 1 through 6 for channel 2 (channels 2 to 4 on the
HP 54601B and HP 54602B).
Service
Verifying Oscilloscope Performance
4–11