section. During testing, the
4
USER
5
and
4
MENU
5
keys are unavailable. After program completion,
these menu keys return to the screen display.
If this program is not currently loaded into RAM and System Diagnostics is to be run from the
front panel and you need information on how to load and run
DOWN_DIAG
, refer to \Running
the System Diagnostics Utilities".
Figure
5-5.
System
Diagnostics
Softkey
Menu
Running Diagnostics Remotely
Use the following procedure to run the REM DIAG utility from an HP 9000 Series 200/300
controller.
1. Load HP BASIC 2.0 with extensions 2.1, or equivalent.
2.
Use
an
HP-IB
cable
to
connect
the
HP-IB
ports
of the computer to the modular spectrum
analyzer
.
This
cable
must
connect
directly
to
the
HP 70004A color display or HP 70001A
mainframe
that
contains
the
master
(
HP
70900B
local
oscillator source LO).
3. Insert the System Diagnostics Software disk into the default drive of the computer.
4.
Type
LOAD "REM_DIAG"
.
5. Press
EXECUTE
.
6. Press
RUN
, then follow the instructions on the computer screen display.
Troubleshooting 5-15