Application
Errors
0001
{
0999
The
numbers
0001|0999
are
reserved
for
error
messages
that are created for user-application
programs
and
loaded
into
RAM.
Error Messages for System Diagnostics
One example of a user-application program is \System Diagnostics," a down-loadable program
(DLP) designed by Hewlett-Packard. The system diagnostics error messages are listed below.
These diagnostic error messages range from 0001 { 0070.
The asterisk (*) printed before some of the following error code messages indicates that the
program stops until the error is no longer being generated. The asterisk does not appear on the
display screen.
Note
When
a
customer-designed
down-loadable program is used, any error numbers
assigned to the program should not duplicate system diagnostics error numbers.
*0001
NO LO?
Cause:
The
program
cannot
nd
an
HP
70900B
local oscillator source at row address 0.
Cure:
Check
and
correct
the
HP-MSIB
switches
on
the HP 70900B local oscillator source.
*0002
HP-MSIB addr 70902
Cause: The HP-MSIB address of the HP 70902A IF section does not follow addressing rules.
Cure: The recommended address of the HP 70902A IF section is in row 2 and the same column
as the HP 70900B local oscillator source.
*0003
HP-MSIB addr 70903
Cause: The HP-MSIB address of the HP 70903A IF section does not follow addressing rules.
Cure: The recommended address of the HP 70903A IF section is in row 4 and the same column
as the HP 70900B local oscillator source
.
*0004
HP-MSIB addr 70904
Cause: The HP-MSIB address of the HP 70904A RF section does not follow addressing rules
.
Cure: The recommended address of the HP 70904A RF section is in row 6 and the same column
as the HP 70900B local oscillator source
.