User
Application
Errors
0001
{
0999
*0012
HP-MSIB addr 70600
Cause: The HP-MSIB address of the Preselector does not follow the addressing rules.
Cure: The recommended address of the Preselector is in row 2 and the column immediately to
the right of the HP 70900B local oscillator source.
*0013
No PRESELECTOR?
Cause: The program cannot nd a Preselector congured to the HP 70900B local oscillator
source.
Cure: Install a Preselector in the system, or correct the HP-MSIB switch settings on the
currently installed preselector.
*0014
Incorrect
System
Cause:
The
system
selected
to
be
tested
requires
modules
which
are
not
present in the system.
Cure:
Select
another
congured
system,
or
recongure
your
system
to
match
the selection.
This
error
will
not
occur
in
Revision
C.00.00
and
later
.
*0015
HP-MSIB addr 70700
Cause: The HP-MSIB address of the HP 70700A digitizer does not follow the addressing rules.
The program can test the HP 70700A digitizer only when it is congured as a slave to the
HP 70900B local oscillator source.
Cure: The recommended address of the HP 70700A digitizer is in row 3 and the same column
as the HP 70900B local oscillator source.
*0020
21.4MHz
input
cable
Cause: The program cannot nd the rear panel 21.4 MHz input signal for the HP 70902A
IF section.
Cure: Check the 21.4 MHz OUT connection on the rear panel of the RF section going to the
21.4 MHz IN connection on the rear panel of the HP 70903A IF section if present, otherwise
the 21.4 MHz IN on the rear panel of the HP 70902A IF section. If this cable is connected
properly, it might be defective.
6-4 Error Messages