0001
{
0999
0033
Log board
Cause: The 70903A IF Section log board signal amplitude is too low.
Cure: Perform the HP 70903A IF section log amplier adjustments. If this does not remedy the
problem, the A2 Log Amplier Assembly requires repair.
0034
Problem/ampl
low?
Cause:
The
HP
70903A
IF section internal 21.4 MHz IF signal amplitude is too low.The
program
stops
testing
the
HP 70903A IF section when this error is generated.
Cure: Perform the HP 70903A IF section module adjustments. It is likely either the A1 LC
wideband lter or A2 log amplier/power supply requires repair.
0035
Step
gain
amplifier
Cause:
The
gain
of
the
step
gain
amplier
of
the
HP
70903A IF section is incorrect.
Cure:
The
step
gain
ampliers
in
the
HP
70902A
IF
section require adjustment or repair.
0036
Res BW-accuracy
Cause: One or more of the HP 70903A IF section resolution bandwidths are not the correct
value.
Cure: The bandwidth lter adjustments in the HP 70903A IF section should be performed. If
this does not remedy the problem, repair the HP 70903A IF section LC bandwidth lters.
0037
Res BW-amplitude
Cause: The amplitude variation between the HP 70903A IF section resolution bandwidths is too
large.
Cure: The bandwidth lter adjustments in the HP 70903A IF section should be performed. If
this does not remedy the problem, repair the HP 70903A IF section LC bandwidth lters
.
0038
Calibration atten
Cause: One or more of the calibration attenuators of the HP 70903A IF section are not
functioning correctly.
Cure: The calibration attenuators in the HP 70903A IF section require adjustment or repair.
Error Messages 6-7