EasyManua.ls Logo

HP 70000 series - Page 316

HP 70000 series
447 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User
Application
Errors
0001
{
0999
0044
Low/high band switch
Cause: The low/high band switch of the RF section is not functioning correctly.
Cure: This error occurs when a signal is detected in one setting of the high-low band coaxial
switch but not the other. This switch is the most likely cause of this error.Test and replace if
necessary.
0045
RF attenuator
Cause: The RF attenuator of the RF section is not functioning correctly.
Cure: Test and replace the RF Attenuator if necessary
0046
Step
gain amplifier
Cause:
The
gain
of the step gain amplier of the HP 70908A RF section is incorrect. This error
occurs
if
the
10 dB step gain amplier in the HP 70908A RF section fails its test.
Cure:
This
amplier is located in the last converter assembly of the HP 70908A RF section. The
last converter assembly should be tested and repaired as required.
0050
Res BW -- CF
Cause: The resolution bandwidth center frequency of one or more of the 70902A IF Sections is
out of adjustment.
Cure:
P
erform
the
HP
70902A
IF
section
Resolution Bandwidth Filter adjustments.
0051
Res
BW
--
CF
ampl
Cause:
The
resolution
bandwidth
of
one
or more of the HP 70902A IF sections is out of
adjustment.
Cure: Perform the HP 70902A IF section Resolution Bandwidth Filter adjustments.
0052
Res BW -- CF ampl
Cause: The resolution bandwidth center frequency of one or more of the HP 70903A
IF sections is out of adjustment.
Cure: Perform the HP 70903A IF section Resolution Bandwidth Filter adjustments.
Error Messages 6-9