0001
{
0999
0064
RF Switch 1
Cause: This RF switch in the HP 70600A preselector or the HP 70601A preselector is not
functioning properly.
Cure: Check A3S1 and replace if required.
0065
RF Switch 2
Cause: This RF switch in the HP 70600A preselector or the HP 70601A preselector is not
functioning properly.
Cure: Check A4S2 and replace if required.
0066
RF Switch 3
Cause:
This
RF
switch
in
the
HP
70600A
preselector
or
the
HP
70601A preselector is not
functioning
properly
.
Cure:
Check
A5S3
and
replace
if
required.
0067
RF
Switch 4
Cause:
This RF switch in the HP 70600A preselector or the HP 70601A preselector is not
functioning
properly.
Cure:
Check A6S4 and replace if required.
0068
RF Switch 1
Cause: The HP 70620 or HP 70621 has failed its switch diagnostic #1. If the preamplier is an
HP 70620A preamplier, S4 or its associated drive circuitry has failed. If the preamplier is an
HP 70620B preamplier or an HP 70621A preamplier, K1 or its associated drive circuitry has
failed.
Cure: Troubleshoot and replace the failed component if necessary.
0069
RF Switch 2
Cause: The HP 70620A preamplier has failed its switch diagnostic #2. This indicates S1 or its
associated drive circuitry has failed. This error applies to the HP 70620A preamplier only.
Cure: Troubleshoot and replace the failed component if necessary.
Error Messages 6-11