Operating
Errors
2000
{
2999
2007
Missing terminator
2008
Output unleveled
2009
Protocol
error
Cause:
Internal
error
due to illegal communication. Due to hardware failure.
Cure: Please document all details possible that lead up to the error and contact your HP
representative
2010
Cmd out of sequence
Cause:
Internal
error
due
to
process
synchronization.
P
ossible
hardware failure.
Cure:
Please document all details possible that lead up to the error and contact your HP
representative.
2011
Memory overflow
Cause: There is not enough available memory for the operation. Examples would be adding a
VARDEF, FUNCDEF,orACTDEF sending AMPCOR data adding new modules to a system that
was almost out of memory because of FUNCDEF's, and so forth
Cure: Some items must be removed from memory. Analyze DLP's to see if there are extra
characters (spaces, and so forth) that could be removed. Expanded memory options for the
Local Oscillator are also available.
2011
A Hardware Selected
Cause: This is notication that a service mode has been enabled. The system is overriding the
automatic conguration and operating as if the hardware
is the \A" version of the HP 70900B
local oscillator source LO module
.
Cure: The power must be cycled to restore the automatic mode.
2012
B Hardware Selected
Cause: This is notication that a service mode has been enabled. The system is overriding the
automatic conguration and operating as if the hardware is the \B" version of the HP 70900B
local oscillator source LO module.
Cure: The power must be cycled to restore the automatic mode.
6-14 Error Messages