Hardware
W
arning
Errors
6000
{6999
Hardware-warning
errors
occur
when
the
hardware
is
faulty
. The spectrum analyzer can still
make
measurements
,
but
the
accuracy
of
the
measurement cannot be guaranteed.
6000
EAROM unprotected
Cause: The memory-enable write switch is set to the WRITE position. It is not set to the
PROTECT position.
Cure: Set the memory-enable write switch to the PROTECT position.
6001
Confidence
test
passed
6002
A6
RAM
checksum
(battery)
6003
FFS won't tune low
6004
FFS
won't
tune
high
6005
Idler tuning range
6006
YTO
tuning
range
6007
MSIB NMAA received
HP-MSIB \no module at address" received
Cause: The module attempted to establish communication with a module at an incorrect
address. The system responded with an NMAA (no module at address).
Cure: If using MSIB communication, verify the address if the module you are trying to
communicate with is correct. Otherwise, document all steps that led to the error condition and
contact your HP representative.
6-22 Error Messages