ASCII:
MISSING 'BEGIN'
STATEMENT
Error Number
193
The citile
you just
downloaded over
the HP-IB
or via
disk
was
not
properly
organized. The
analyzer is
unable to
read the
\BEGIN" statement.
ASCII:
MISSING
'CITIFILE'
STATEMENT
Error
Number
194
The
citile
you
just
downloaded
over
the
HP-IB
or
via
disk
was
not properly
organized.
The
analyzer
is
unable
to
read
the
\CITIFILE"
statement.
ASCII:
MISSING
'DATA'
STATEMENT
Error
Number
195
The
citile you
just
downloaded
over
the
HP-IB
or
via
disk
was
not
properly
organized.
The
analyzer
is
unable
to read
the
\D
A
T
A"
statement.
ASCII:
MISSING
'VAR'
STATEMENT
Error
Number
196
The
citile
you
just
downloaded
over
the
HP-IB
or
via
disk
was not
properly
organized.
The
analyzer
is
unable
to
read
the
\V
AR"
statement.
AVERAGING
INVALID
ON
NON-RATIO
MEASURE
Error
Number
13
Y
ou
cannot
use
sweep-to-sweep
averaging
in
single-input
measurements
.
Sweep-sweep
averaging
is
valid
only
for
ratioed
measurements
(A/R,
B/R,
A/B,
and
S-parameters).
Y
ou
can
use
noise
reduction
techniques,
such as
narrower
IF bandwidth,
for single
input measurements
.
BAD
FREQ
FOR
HARMONIC
OR
FREQ
OFFSET
Error
Number
181
Y
ou
turned
on
time
domain
or
recalled
a
calibration that
resulted
in
start
and
stop frequencies that are beyond the allowable limits
.
BATTERY FAILED. STATE MEMORY CLEARED
Error Number
183
The battery protection of the non-volatile CMOS memory has failed. The CMOS
memory has been cleared. Refer to the
HP 8753D Network Analyzer Service
Guide
for battery replacement instructions. See Chapter 12, \Preset State and
Memory Allocation," for more information about the CMOS memory.