OUTPUT COMPLETED
Information
Message
The analyzer
has completed
outputting data
to the
printer or
plotter
.
The
analyzer can
now accept
another copy
command.
CORRECTION
AND
DOMAIN
RESET
Error
Number
65
When
you
change
the
frequency
range
,
sweep
type
,
or
number
of points
,
error-correction
is
switched
o
and
the
time
domain
transform
is
recalculated,
without
error-correction.
Y
ou
can
either
correct
the
frequency
range
,
sweep
type
,
or
number
of
points
to
match
the
calibration,
or
perform
a new
calibration.
Then
perform
a
new
time
domain
transform.
CORRECTION
CONSTANTS
NOT
STORED
Error
Number
3
A
store
operation
to
the
EEPROM
was not
successful. Y
ou must
change
the
position
of
the
jumper
on
the
A9 CPU
assembly.
Refer to
the \A9
CC
Jumper
P
osition
Procedure"
in
the
\A
djustments
and
Correction Constants"
chapter of
the
HP
8753D
Network
Analyzer
Service
Guide
.
CORRECTION
TURNED
OFF
Error
Number
66
Critical
parameters
in
your
current
instrument
state
do not
match
the
parameters
for
the
calibration
set,
therefore
correction
has been
turned
o.
The
critical
instrument
state
parameters
are
sweep
type,
start frequency
,
frequency
span,
and
number
of
points
.
CURRENT
PARAMETER
NOT
IN
CAL
SET
Error
Number
64
Correction
is
not valid
for your
selected measurement
parameter.
Either
change the
measurement parameters
or perform
anew
calibration.
D2/D1 INVALID WITH SINGLE CHANNEL
Error Number
130
You
can only make a D2/D1 measurement if both channels are on.
10-6 Error Messages