Loading...
DOS
NAME LIMITED
TO
8
CHARS
+
3
CHAR
EXTENSION
Error Number
180
ADOS
le name
must meet
the following
criteria:
minimum of
1 character
format is
filename.ext
maximum of
8 characters
in
the
lename
maximum of
3 characters
in
the
extension
eld
(optional)
a dot
separates the
lename
from
the
extension
eld
(the
dot
is
not
part
of
the name
on the
disk)
DUPLICATING
TO
THIS
SEQUENCE
NOT
ALLOWED
Error
Number
125
A
sequence
cannot
be
duplicated
to
itself.
EXCEEDED
7
STANDARDS
PER
CLASS
Error
Number
72
When
modifying
calibration
kits
,
you
can
dene
a maximum
of seven
standards
for
any
class
.
EXTERNAL
SOURCE
MODE
REQUIRES
CW
TIME
Error
Number
148
An
external
source
can
only
be phase
locked and
measured
in
the
CW
time
sweep
mode
.
EXT
SOURCE
NOT
READY
FOR
TRIGGER
Error Number
191
There is
a hardware
problem with
the HP
8625A external
source.
V
erify
the
connections
between
the
analyzer
and
the
external
source
.
If
the connections
are correct, refer to the source operating manual.
EXT SRC: NOT ON/CONNECTED OR WRONG
ADDR
Error Number
162
The analyzer is unable to communicate with the external source
. Check the
connections and the HP-IB address on the source
.
Error Messages 10-9