PWR
MTR NOT
ON/CONNECTED
OR
WRONG
ADDRS
Error
Number
117
The
power
meter
cannot
be
accessed
by
the
analyzer
.
V
erify
that
the
power
meter
address
and
model
number
set
in
the
analyzer
match
the
address
and
model
number of
the
actual
power
meter
.
REQUESTED DATA
NOT
CURRENTLY
AVAILABLE
Error
Number
30
The
analyzer
does not
currently contain
the data
you have
requested.
F
or
example
,
this condition
occurs when
you request
error term
arrays
and
no
calibration
is
active.
SAVE FAILED.
INSUFFICIENT MEMORY
Error
Number
151
Y
ou
cannot
store
an
instrument
state
in
an
internal
register
due
to
insucient
memory
.
Increase
the
available
memory by
clearing one
or more
save/recall
registers
and
pressing
4
PRESET
5
,
or
by
storing
les to
a disk.
SELECTED
SEQUENCE
IS
EMPTY
Error
Number
124
The
sequence
you
attempted
to
run
does
not
contain
instrument
commands
.
SELF
TEST
#n
FAILED
Service
Error
Number 112
Internal
test
#n
has
failed.
Several
internal test
routines are
executed at
instrument preset.
The analyzer
reports the
rst
failure
detected.
Refer
to
the
HP 8753D
Network Analyzer
Service Guide
for
troubleshooting
information
on
internal tests
and the
self-diagnose feature
.
SEQUENCE ABORTED
Error Number
157
The sequence running was stopped prematurely when you pressed the
4
LOCAL
5
key.
SEQUENCE MAY HAVE CHANGED, CAN'T CONTINUE
Error Number
153
When you pause a sequence, you cannot continue it if you have modied it.
You must start the sequence again.
Error Messages 10-21