-222,"Data out of range;
DC
Offset is out of range"
Triggered
by:
V0LT:OFFS command. Requested
DC
offset voltage
is out of absolute bounds.
-222,"Data out of range; Dwell time of
LOG
interval too low (<O.ls)"
Triggered by: FREQ:MODE command. Dwell time for one of the
LOG intervals is less than 0.1 sec.
-222,"Data out of range; Frequency is too low"
Triggered by: FREQ command. FIX frequency is
<
0.
-222,"Data out of range; Frequency too high for output function"
Triggered by: FUNC or FREQ command. FREQ is too high for the
output function.
-222,"Data out of range; Marker index is out range"
Triggered by: MARK:STAT and MARK:FREQ command. Attempt
to set marker index outside the range
1
..
9.
-222,"Data out of range; Phase offset is out of range (-720
..
720 deg)"
Triggered by: PHAS command. Attempt to set phase outside the
range -720
..
720 degree.
-222,"Data out of range; *SAV/*RCL register number out of range (0
..
9)''
Triggered by: *SAV or "RCL command. Command argument
(register number) is outside the range 0
..
9.
-222,"Data out of range; Start frequency in list is too high or negative"
Triggered by: FREQ:MODE command. One of the sequenced start
frequencies in the list, is too high or negative.
-222,"Data out of range; Start frequency is too low"
Triggered by: FREQ:MODE command. FREQ:STAR value is too
low (negative).
-222,"Data out of range; Start frequency too high for output function"
Triggered
by:
FREQMODE
command.
FREQSTAR
value
is
too
high for the output function.
-222,"Data out of range; Stop frequency in list is too high or negative"
Triggered by: FREQ:MODE command. One of the sequenced stop
frequencies in the list, is too high or negative.
-222,"Data out of range; Stop frequency is too low"
Triggered by: FREQ:MODE command. FREQ:STOP value is too
low (negative).
-222,"Data out of range; Stop frequency too high for output function"
Triggered by: FREQ:MODE command. FREQ:STOP value is too
high for the output function.
Error
Messages
E-5