[SOURce]
This command defines the frequency characteristics
of
an output
:FREQuency
signal.
Subsystem Syntax
[:SOURcel
:FREQuency
[:CW
1
FIXed] <numeric value>
:MODE CW
I
FIXed
(
SWEep
I
LIST
:
STARt <numeric value>
:STOP <numeric value>
:
CENTer <numeric value>
:SPAN <numeric value>
:HOLD <boolean>
:LINK
CENTer
I
STARt
(
STOP
:FULL <event
>
:FREQuency[:CW
I
This command sets non-swept frequency of the instrument output
:FIX&]
signal. The value is specified by a digital number representing
Hz
i.e.
10000
specifies
10kHz.
:FREQuency:MODE
Determines which set of commands control the frequency subsystem.
The settings have the following meanings:
rn
CW
I
FIXed: the frequency is determined by :FREQ[:CW] or
:FREQ[:FIX].
rn
Sweep: The source is in swept mode and frequency is determined
by :FREQ:STAR, :STOP, :CENT and :SPAN
rn
LIST: The source is in list sequence mode and frequency is
determined by the :LIST:FREQ command.
:FREQuency :STARt
These two commands set the start and stop frequencies for a sweep.
:STOP
Changing the start frequency will change the centre and span settings
but not the stop.