EasyManua.ls Logo

HT COMBI419 - Page 67

Default Icon
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
400 Series
EN - 66
Environmental parameters (AUX)
Parameter Range Resolution
Transduced
signal
Uncertainty
Temperature
-20.0 80.0°C
0.1°C
-20 +80mV
(2.0rdg + 2dgt)
-4.0 176.0°F
0.1°F
-4 +176mV
Humidity
0.0 100.0% RH
0.1% RH
0 +100mV
DC voltage
(0.0 999.9mV)
0.1mV
(0.2 999.9mV)
Luminance
0.001 20.00Lux 0.001 0.02Lux 0 +100mV
0.1 2000Lux 0.1 2Lux 0 +100mV
1 20000Lux 0.1 2Lux 0 +100mV
Mains analisys (PWR)
Frequency measurement
Range [Hz] Resolution [Hz] Uncertainty
47.0 63.0
0.1
(2.0rdg + 2dgt)
Voltage range 5.0 265.0
Current range 0.005 1.2 x FS
Voltage measurement
Range [V] Resolution [V] Uncertainty
5.0 265.0
0.1V
(0.5rdg + 2dgt)
Max crest factor 1,5
Frequency 47.0 63.0 Hz
Voltage harmonics measurement
Range [V] Resolution [V] Order Uncertainty
0.0 265.0
0.1V
2 15 (2.0rdg + 5dgt)
16 49 (5.0rdg + 10dgt)
Frequency fundamental 47.0 63.0 Hz
Current measurement
Range [A] Resolution [A] Uncertainty
0.005 1.2 x FS
See Tab. 2
(1.0rdg + 2dgt)
Max crest factor 3
Frequency 47.0 63.0 Hz
Current harmonics measurement
Range [V] Resolution [A] Order Uncertainty
0.005 1.2 x FS
See Tab. 2
2 15 (2.0rdg + 5dgt)
16 49 (5.0rdg + 10dgt)
Frequency fundamental 47.0 63.0 Hz
Current fundamental 0.020 x FS
Full scale [A] Resolution [A] Full scale [A] Resolution [A]
1 0.001 300 0.1
10 0.01 400 0.1
30 0.01 1000 1
100 0.1 2000 1
200 0.1 3000 1
Tab. 2: Clamp and instrument full scale and resolution values

Table of Contents

Related product manuals