3
DDoo
nnoott
use Dopplers in the presence of
flammable gases such as anaesthetic
agents.
DDoo
nnoott
use in the sterile field unless additional
barrier precautions are taken.
DDoo
NNoott
:
Immerse in any liquid, (Except
FFDD11//
FFDD33
probe)
use solvent cleaner,
use high temperature sterilising
processes (such as autoclaving),
use E-beam or gamma radiation
sterilisation.
This product contains sensitive electronics,
therefore, strong radio frequency fields could
possibly interfere with it. This will be indicated
by unusual sounds from the loudspeaker. We
recommend that the source of interference is
identified and eliminated.
DDoo
nnoott
use on the eye.
If outputs 1 and 2 are used simultaneously,
the system should comply with EN60601-1-1.
Any equipment connected to outputs 1 or 2
should comply with EN60601-1, EN60950,
EN60065, EN60335 or EN61010.
DDoo
nnoott
dispose of batteries in fire as this can
cause them to explode.
DDoo
nnoott
attempt to recharge normal dry-cell
batteries. They may leak, cause a fire or even
explode.
The High Sensitivity Dopplers are screening
tools to aid the healthcare professional and
should not be used in place of normal
vascular or fetal monitoring. If there is doubt
as to vascularity or fetal well-being after
using the unit, further investigations should be
undertaken immediately using alternative
techniques.
We recommend that exposure to ultrasound
should be kept As Low As Reasonably
Achievable - (ALARA guidelines). This is
considered to be good practice and should
be observed at all times.
The main unit is not waterproof and must not
be immersed. For underwater use where
contamination or cross-infection may occur,
additional barrier precautions must be taken.
WWAARRNNIINNGG//CCAAUUTTIIOONNSS
&&
SSAAFFEETTYY
Refer to diagram on inside front cover for
Doppler Measuring sites and
Recommended Probes.
DDuurriinngg
UUssee
An automatic noise reduction feature
operates on low level signals to improve
sound quality.
To replace battery see diagram.
AAfftteerr
UUssee
1. Press and release the On/Off
button. If you forget to switch the
unit off, it will automatically shut-off
after 3 minutes.
2. Refer to the cleaning section
before storing or using the unit on
another patient.
3. Store unit together with probe and
accessories in the soft carry case
provided.
The control unit and the body of the probe
are robust and require no special handling.
However, the probe tip is delicate and must
be handled with care.
CClleeaanniinngg
Remove excess gel before parking the
probe. Clean with a damp cloth impregnated
with mild detergent. Do not allow fluid to
seep into either unit.
For disinfection use a soft cloth with Sodium
Hypochlorite 1000ppm or alcohol.
Alcohol saturated swabs can be used on the
probe tip.
Please be sure to check your local control of
infection policies, or any equipment cleaning
procedures.
Phenolic, detergent based disinfectants
containing cationic surfactants, ammonia
based compounds, or antiseptic solutions
such as Steriscol or Hibiscrub should never
be used on any part of the system.
CCoouupplliinngg
GGeell
Use water based ultrasound gel ONLY.
FFDD11//
FFDD33
Gel not required when probe used
underwater.
GGEENNEERRAALL
OOPPEERRAATTIIOONN
&&
MMAAIINNTTEENNAANNCCEE
Means:
Attention
consult
this
manual.
Refer
to
safety
section.