How to fix mattress cells that will not inflate on Huntleigh Nimbus 3 Professional Medical Equipment?
- TTiffany HicksAug 1, 2025
The vent valves may be open. Close Vent Valves.

How to fix mattress cells that will not inflate on Huntleigh Nimbus 3 Professional Medical Equipment?
The vent valves may be open. Close Vent Valves.
What to do if Huntleigh Medical Equipment shows flashing POWER and symbol?
This indicates a Power Fail Alarm, meaning the pump has detected that mains power has been removed. Re-apply mains power or switch off the pump using the POWER switch on the control panel. If power failure is prolonged, switch to TRANSPORT mode and disconnect the tubeset. The mattress will remain inflated for 12 hours.
What to do if Huntleigh Medical Equipment displays LOW PRESSURE?
This may happen due to several reasons. First, check the tubeset connectors and make sure they are securely connected to the pump and mattress. Second, ensure the CPR control is fully closed. Third, verify that the Transport control on the mattress is set to NORMAL. If none of these resolve the issue, there may be a leak in the system, and you should call the service engineer.
What to do if Huntleigh Medical Equipment displays flashing PUMP FAULT and symbol?
This indicates an internal pump malfunction. Call the service engineer.
What to do if Huntleigh Nimbus 3 displays HIGH PRESSURE?
This may be because the tubeset is blocked or the AutoMatt sensor pad is blocked. Check that the tubeset is not kinked and that the AutoMatt sensor pad is flat and not kinked.
Why Huntleigh Medical Equipment shows LOW PRESSURE and WAIT?
This may happen because the pump is inflating the mattress, or the CPR control is not fully closed. Both indicators will extinguish when the operating pressure is reached. Close CPR control.
Why Huntleigh Medical Equipment displays symbol?
This indicates that the pump needs a service. Call the service engineer.
| Type | Alternating Pressure Mattress System |
|---|---|
| Usage | Prevention and treatment of pressure ulcers |
| Pump Type | Digital pump |
| Warranty | 2 years |
| Pressure Range | Adjustable |
| Pressure Settings | Adjustable |
| Cycle Time | Typically 10 minutes |
| Modes | Alternating, Static |
| Mattress Material | Nylon/PU |
| Mattress Construction | Cellular |
| Control Unit | Digital control unit with user-friendly interface |
| Power Supply | 220-240V, 50Hz |
| Pump Unit Dimensions | 29 cm x 19 cm x 12 cm |
| Noise Level | <30 dB (quiet operation) |
Important warnings regarding suffocation risk, electrical hazards, and flammable gases.
Outlines advanced features like AutoMatt, CPR, Transport, and Tubeset for the Professional model.
Guides on unpacking and physically installing the mattress onto the bed frame.
Describes pump controls, indicators, and operational modes on the front panel.
Step-by-step instructions to activate and reset the CPR control for rapid deflation.
A table listing common indicators, possible causes, and remedies for system malfunctions.